CAS I/II Absorbers that are parts of the BleaseSirius Anesthesia Workstation

Recall Z-0352-2014, part of FDA recall event 66721. See Del Mar Reynolds Medical's company page.

Del Mar Reynolds Medical, Ltd. started this Class I device recall on Oct 17, 2013 (voluntary: firm initiated); FDA lists it as terminated, terminated Apr 24, 2014.

  • Classification: Class I
  • Started: Oct 17, 2013
  • Status: Terminated

Why It Was Recalled

Class I

Spacelabs Healthcare Ltd. of United Kingdom recalls , and Service Kits, Part Number 050-0659-00 and 050-0901-00. The Bag to-Vent switch in CAS I/II Absorbers may fail due to loose fastening hardware in an absorber.

Record

Product
CAS I/II Absorbers that are parts of the BleaseSirius Anesthesia Workstation, the BleaseFocus Anesthesia Workstation with the following part numbers: 12200900; 12200901; 12200902; 12200903; 14200100, and service kits part number 050-0659-00 and 050-0901-00. BleaseFocus Anesthesia Workstation: The Blease Frontline Plus Range, Anesthesia Machines are intended for use in the hospital environment and locations not requiring portability. It may be used for the delivery of oxygen, air and nitrous oxide in a controlled manner to various patient breathing circuits with or without the use of a mechanical ventilator, and may be used for the delivery of anesthetic vapor by use of a dismountable vaporizer. This device is intended for use only by a suitably qualified physician. BleaseSirius Anesthesia Workstation: The Spacelabs BleaseSirius Anesthesia Workstation is intended for use in the hospital environment and operating room. It may be used for the delivery of oxygen, air and nitrous oxide in a controlled manner to various patient breathing circuits with or without the use of mechanical ventilator, and may be used for the delivery of anesthetic vapor by use of a dismountable vaporizer. The device is intended for use only by a suitably qualified physician.
Quantity
37 units (19 in the US and 18 International)
Lots and codes
SERIAL NUMBERS IN THE US: ABSO-001754, ABSO-004081, ABSO-100111, ABSO-100112, ABSO-100113, ABSO-100114, ABSO-100117, ABSO-100120, ABSO-100121, ABSO-100122, ABSO-100123, ABSO-100124, ABSO-100125, ABSO-100126, ABSO-100127, ABSO-100128, ABSO-100132, ABSO-100134, and ABSO-100135. SERIAL NUMBERS INTERNATIONAL: ABSO-100098, ABSO-100099, abso-100100, ABSO-100101, abso-100102, ABSO-100103, abso-100104, abso-100105, abso-100106, ABSO-100107, ABSO-100108, ABSO-100109, ABSO-100110, ABSO-100129, ABSO-100130, ABSO-100131, ABSO-100136, AND ABSO-100137.
Distribution
Worldwide Distribution - USA (nationwide) including Oklahoma and North Carolina and Internationally to Libyan Arab Jamahiriya, Finland, Jordan, Mexico, Papua New Guinea, and Taiwan.
Recalling firm
Del Mar Reynolds Medical, Ltd., Hertford, United Kingdom
How the firm notified
Telephone
FDA classified
Nov 20, 2013