EDS 3 CSF External Drainage System
Recall Z-0635-2024, part of FDA recall event 93550. See Natus Medical's company page.
Natus Medical Incorporated started this Class II device recall on Nov 28, 2023 (voluntary: firm initiated); FDA lists it as ongoing.
- Classification: Class II
- Started: Nov 28, 2023
- Status: Ongoing
- Products in this event: 2
Why It Was Recalled
Class IISterility assurance cannot be guaranteed for external drainage systems due to incomplete bioburden testing.
Record
- Product
- EDS 3 CSF External Drainage System, REF: NT821731C, STERILEEO, Rx only
- Quantity
- 3,336 units
- Lots and codes
- UDI: 00382830050555/ Lot #: 1FX0004588937 1FX0004588938 1FX0004588939 1FX0004663828 1FX0004663825 1FX0004663826 1FX0004663824 1FX0004663827 1FX0004857869 1FX0004857871 1FX0004663829 1FX0004663830 1FX0004857870 1FX0004763454 1FX0004833126 1FX0004833140 1FX0004833139 1FX0004857872 1FX0004857873 1FX0004833127 1FX0004833136 1FX0004763455 1FX0004763460 1FX0004763463 1FX0004763464 1FX0004763462 1FX0004763461 1FX0004763459 1FX0004857877 1FX0004857875 1FX0004857874 1FX0004763457 1FX0004857876 1FX0004763465 1FX0004763456 1FX0004833141 1FX0004763458 1FX0004857882 1FX0004857881 1FX0004857880…
- Distribution
- US Nationwide distribution including in the states of AL, AR, AZ, CA, CO, CT, DC, FL, GA, IL, IN, KS, KY, LA, MA, MD, ME, MI, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NY, OH, OK, OR, PA, RI, SC, TN, TX, UT, VA, WA, WI, WV.
- Recalling firm
- Natus Medical Incorporated, San Diego, CA, United States
- How the firm notified
- Letter
- FDA classified
- Jan 4, 2024
Other Products in This Recall
Recall event 93550- External Drainage System Collection Bag, REF: NT821732C, STERILEEO, Rx onlyZ-0634-2024 · Class II · Nov 28, 2023 · Ongoing
Natus Medical Incorporated's Other Recalls
Newest first, one product per recall event; search them all- The Natus Cranial Access Kit is designed to be a complete, single use and disposable accessory and…Z-0735-2023 · Class II · Nov 18, 2022 · Ongoing
- The Olympic Pasteurmatic 3000 and Olympic Pasteurmatic 3500 are intended for cleaning and high-level…Z-2701-2017 · Class II · May 22, 2017 · Terminated
- Second generation neoBLUE LED Phototherapy Light (neoBLUE 2) sold between 2003 and 2008. Catalog Part Number…Z-1924-2017 · Class II · Apr 3, 2017 · Terminated
- Natus neoBLUE blanket LED Phototherapy system, neonatal phototherapy. Used to provide treatment for neonatal…Z-0770-2017 · Class II · Oct 28, 2016 · Terminated
- Updated Natus neoBLUE blanket LED Phototherapy Systems. It provides phototherapy underneath the infant and…Z-0769-2017 · Class II · Sep 16, 2016 · Terminated
- The Medix NatalCare LX Intensive Care Incubator. Neonatal incubator. Item numbers: DMNCLX1; KITNC-STLX0003;…Z-0056-2017 · Class II · Aug 25, 2016 · Terminated
- Replacement HEPA filter for Natus Sterile-Drier Model 43 and Model 44. Filter Catalogue Part Number: 54343,…Z-1878-2016 · Class II · Apr 22, 2016 · Terminated
- Natus neoBLUE Phototherapy system, Catalogue Part Number: 010066. Top Assembly Part Numbers: 001103, 001314,…Z-2001-2016 · Class II · Apr 11, 2016 · Terminated
FDA's weekly Enforcement Report, as openFDA publishes it. A recall is classified by how likely the product is to cause harm: Class I is the most serious. openFDA device enforcement reports