Natus neoBLUE Phototherapy system
Recall Z-2001-2016, part of FDA recall event 73169. See Natus Medical's company page.
Natus Medical Incorporated started this Class II device recall on Apr 11, 2016 (voluntary: firm initiated); FDA lists it as terminated, terminated Dec 14, 2016.
- Classification: Class II
- Started: Apr 11, 2016
- Status: Terminated
The firm is adding the warning statement to the Instruction for Use for the neoBLUE LED Phototherapy System and provide a laminated Quick Guide to affix to each device.
- Product
- Natus neoBLUE Phototherapy system, Catalogue Part Number: 010066. Top Assembly Part Numbers: 001103, 001314, 002110, 040906, 040904, 040756, 040758, 040658, 040680. Product Usage: Used to provide treatment for neonatal hyperbilirubinemia
- Quantity
- 13541 units total (7840 units in the US and 5701 units outside the US)
- Lots and codes
- Serial numbers of units distributed in American Samoa, Guam, Puerto Rico and in the US: .22938, .23282, .23283, .23289, .23291, .23302, .23307, .23471, .23473, .23478, .23479, .23480, .23481, .23483, .23538, .23697, .23708, .23887, .23888, .23895, .23916, .23939, .23968, .23969, .23972, .23983, .23993, .23994, .23998, 040756-1, 040756-2, 1, 10090, 10507, 10559, 10592, 10595, 10597, 10598, 10613, 10633, 10634, 10635, 10636, 10637, 10638, 10639, 10640, 10642, 10643, 10645, 10646, 10659, 10660, 10661, 10662, 10663, 10666, 10667, 10668, 10671, 10678, 10679, 10680, 10681, 10682, 10683, 10684,…
- Distribution
- Worldwide Distributed - US Nationwide in the US including American Samoa, Guam Puerto Rico and in the following countries: Argentina, Australia, Austria, Bangladesh, Belgium, Bermuda, Bolivia, Brunei Darussalam, Canada, China, Colombia, Costa Rica, Croatia, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, Finland, France, Germany, Greece, Guatemala, Hong Kong, Hungary, India, Indonesia, Iraq, Ireland, Israel, Italy, Japan, Jordan, , Kenya, , Kuwait, Latvia, Lithuania, Malaysia, Moldova Republic of, Morocco, Namibia, Netherlands, Norway, Oman, Pakistan, Panama, Paraguay, Peru, Philippines, Poland, Portugal, Qatar, Romania, Russia, Saudi Arabia, Serbia, Singapore, Slovakia (Slovak Republic), Slovenia, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, United Arab Emirates, United Kingdom, Uruguay, Viet Nam, Virgin Islands (British), and Yemen.
- Recalling firm
- Natus Medical Incorporated, Seattle, WA, United States
- How the firm notified
- Letter
- FDA classified
- Jun 15, 2016
Newest first, one product per recall event; search them all
- External Drainage System Collection Bag, REF: NT821732C, STERILEEO, Rx onlyZ-0634-2024 · Class II · Nov 28, 2023 · Ongoing
- The Natus Cranial Access Kit is designed to be a complete, single use and disposable accessory and instrumentation set…Z-0735-2023 · Class II · Nov 18, 2022 · Ongoing
- The Olympic Pasteurmatic 3000 and Olympic Pasteurmatic 3500 are intended for cleaning and high-level disinfection of…Z-2701-2017 · Class II · May 22, 2017 · Terminated
- Second generation neoBLUE LED Phototherapy Light (neoBLUE 2) sold between 2003 and 2008. Catalog Part Number 010066. …Z-1924-2017 · Class II · Apr 3, 2017 · Terminated
- Natus neoBLUE blanket LED Phototherapy system, neonatal phototherapy. Used to provide treatment for neonatal…Z-0770-2017 · Class II · Oct 28, 2016 · Terminated
- Updated Natus neoBLUE blanket LED Phototherapy Systems. It provides phototherapy underneath the infant and can be used…Z-0769-2017 · Class II · Sep 16, 2016 · Terminated
- The Medix NatalCare LX Intensive Care Incubator. Neonatal incubator. Item numbers: DMNCLX1; KITNC-STLX0003;…Z-0056-2017 · Class II · Aug 25, 2016 · Terminated
- Replacement HEPA filter for Natus Sterile-Drier Model 43 and Model 44. Filter Catalogue Part Number: 54343, 54344, and…Z-1878-2016 · Class II · Apr 22, 2016 · Terminated
FDA's weekly Enforcement Report, as openFDA publishes it. A recall is classified by how likely the product is to cause harm: Class I is the most serious. openFDA device enforcement reports