Updated Natus neoBLUE blanket LED Phototherapy Systems. It provides phototherapy…
Recall Z-0769-2017, part of FDA recall event 75228. See Natus Medical's company page.
Natus Medical Incorporated started this Class II device recall on Sep 16, 2016 (voluntary: firm initiated); FDA lists it as terminated, terminated Sep 1, 2017.
- Classification: Class II
- Started: Sep 16, 2016
- Status: Terminated
Why It Was Recalled
Class IIThe firm sent the Technical Bulletin to consignees who received the reliability updated neoBLUE blanket LED Phototherapy Systems between March 2016 to August 2016. The reliability update to prevent the early failure of the fiberoptic pads experienced by some customers on older systems. These failures on previous systems involved discoloration and melting of the fiberoptic bundle at the connector that is inserted in the neoBLUE blanket light box.
Record
- Product
- Updated Natus neoBLUE blanket LED Phototherapy Systems. It provides phototherapy underneath the infant and can be used in a bassinet, open bed, radiant warmer, incubator, or while holding the infant. Affected systems including: 1) Updated light boxes (serial number xxxx004283 or greater) have protection built in to turn off the device if an overheat condition is detected. 2) Updated fiberoptic blankets (Lot Nxxxx16-xx or greater) have an improved fiberoptic cable end to reduce the degradation and increase its useful life. Affected Part Numbers and Description: 006244 neoBLUE blanket LED Phototherapy Systems with Large blanket/pad 006254 neoBLUE blanket LED Light Box 006895 neoBLUE blanket LED Phototherapy Systems with Small blanket/pad 007299 neoBLUE blanket LED Phototherapy Systems with Large blanket/pad EUR 007300 neoBLUE blanket LED Phototherapy Systems with Small blanket/pad EUR 007296 neoBLUE blanket LED Phototherapy Systems with Large blanket/pad UK 007298 neoBLUE blanket LED Phototherapy Systems with Small blanket/pad UK 007301 neoBLUE blanket LED Phototherapy Systems with Large blanket/pad AUS 007302 neoBLUE blanket LED Phototherapy Systems with Small blanket/pad AUS
- Quantity
- 713 units total (403 units in US and 310 units outside US)
- Lots and codes
- *************Serial Numbers of Units distributed in the US ************** 004329, 004330, 004349, 004360, 004361, 004362, 004363, 004364, 004365, 004366, 004367, 004369, 004370, 004371, 004417, 004421, 004426, 004436, 004437, 004443, 004448, 004449, 004454, 004456, 004473, 004481, 004488, 004490, 004494, 004495, 004502, 004507, 004517, 004523, 004525, 004526, 004527, 004528, 004530, 004533, 004539, 004541, 004544, 004545, 004546, 004548, 004552, 004553, 004554, 004555, 004556, 004557, 004558, 004567, 004568, 004573, 004578, 004581, 004582, 004584, 004588, 004591, 004602, 004608, 004630,…
- Distribution
- Worldwide Distribution - USA (nationwide) Distribution to the states of : AL, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MS, NC, ND, NJ, NM, NY, OH, OK, OR, PA, SC, TN, TX, UT, VA, WI, WV and Puerto Rico. , and to the countries of : Argentina, Austria, Canada, Chile, China, Denmark, Dominican Republic, Finland, France, Germany, Hong Kong, Hungary, Indonesia, Italy, Latvia, Malaysia, Namibia, Paraguay, Poland, Qatar, Romania, Russia, Saudi Arabia, Singapore, Slovenia, South Africa, Spain, Taiwan, Trinidad and Tobago, Turkey, United Arab Emirates, United Kingdom, Uruguay, Viet Nam, and Virgin Islands (British).
- Recalling firm
- Natus Medical Incorporated, Seattle, WA, United States
- How the firm notified
- Letter
- FDA classified
- Dec 15, 2016
Natus Medical Incorporated's Other Recalls
Newest first, one product per recall event; search them all- External Drainage System Collection Bag, REF: NT821732C, STERILEEO, Rx onlyZ-0634-2024 · Class II · Nov 28, 2023 · Ongoing
- The Natus Cranial Access Kit is designed to be a complete, single use and disposable accessory and…Z-0735-2023 · Class II · Nov 18, 2022 · Ongoing
- The Olympic Pasteurmatic 3000 and Olympic Pasteurmatic 3500 are intended for cleaning and high-level…Z-2701-2017 · Class II · May 22, 2017 · Terminated
- Second generation neoBLUE LED Phototherapy Light (neoBLUE 2) sold between 2003 and 2008. Catalog Part Number…Z-1924-2017 · Class II · Apr 3, 2017 · Terminated
- Natus neoBLUE blanket LED Phototherapy system, neonatal phototherapy. Used to provide treatment for neonatal…Z-0770-2017 · Class II · Oct 28, 2016 · Terminated
- The Medix NatalCare LX Intensive Care Incubator. Neonatal incubator. Item numbers: DMNCLX1; KITNC-STLX0003;…Z-0056-2017 · Class II · Aug 25, 2016 · Terminated
- Replacement HEPA filter for Natus Sterile-Drier Model 43 and Model 44. Filter Catalogue Part Number: 54343,…Z-1878-2016 · Class II · Apr 22, 2016 · Terminated