GE Healthcare Carescape Central Station (CSCS) V3 MAS800 Desktop with software version…

Recall Z-0214-2026, part of FDA recall event 97646. See Wipro GE Healthcare's company page.

Wipro GE Healthcare Private Ltd. started this Class II device recall on Sep 5, 2025 (voluntary: firm initiated); FDA lists it as ongoing.

  • Classification: Class II
  • Started: Sep 5, 2025
  • Status: Ongoing
  • Products in this event: 2

Why It Was Recalled

Class II

Carescape Central Station (CSCS), software version V3.0.5, can cause the system to enter a continuous reboot cycle. If this occurs, a loss of central monitoring of connected patients could result. This issue occurs only when the number of central stations connected to the network is 118, 119 or 120.

Record

Product
GE Healthcare Carescape Central Station (CSCS) V3 MAS800 Desktop with software version V3.0.5, Part Number 5867474-03; Medical Display, cathode ray tube
Quantity
2365 units
Lots and codes
GTIN 00195278343383, Serial Numbers: STV25225456SA, STV25225455SA, STV25266464SA, STV25235817SA, STV25235816SA, STV25286682SA, STV25225594SA, STV25235792SA, STV25235774SA, STV25225550SA, STV25235626SA, STV25235625SA, STV25184592SA, STV25235628SA, STV25235631SA, STV25184595SA, STV25235629SA, STV25235632SA, STV25235630SA, STV25276515SA, STV25225510SA, STV25276516SA, STV25225505SA, STV25225508SA, STV25225509SA, STV25225507SA, STV25225506SA, STV25113369SA, STV25204986SA, STV25205160SA, STV25205151SA, STV25205161SA, STV25062736SA, STV25256133SA, STV25225512SA, STV25296702SA, STV25296701SA,…
Distribution
Worldwide distribution.
Recalling firm
Wipro GE Healthcare Private Ltd., Bengaluru, India
How the firm notified
Letter
FDA classified
Oct 21, 2025

Other Products in This Recall