GE Healthcare Carescape Central Station (CSCS) V3 MAS800 Desktop with software version…
Recall Z-0214-2026, part of FDA recall event 97646. See Wipro GE Healthcare's company page.
Wipro GE Healthcare Private Ltd. started this Class II device recall on Sep 5, 2025 (voluntary: firm initiated); FDA lists it as ongoing.
- Classification: Class II
- Started: Sep 5, 2025
- Status: Ongoing
- Products in this event: 2
Carescape Central Station (CSCS), software version V3.0.5, can cause the system to enter a continuous reboot cycle. If this occurs, a loss of central monitoring of connected patients could result. This issue occurs only when the number of central stations connected to the network is 118, 119 or 120.
- Product
- GE Healthcare Carescape Central Station (CSCS) V3 MAS800 Desktop with software version V3.0.5, Part Number 5867474-03; Medical Display, cathode ray tube
- Quantity
- 2365 units
- Lots and codes
- GTIN 00195278343383, Serial Numbers: STV25225456SA, STV25225455SA, STV25266464SA, STV25235817SA, STV25235816SA, STV25286682SA, STV25225594SA, STV25235792SA, STV25235774SA, STV25225550SA, STV25235626SA, STV25235625SA, STV25184592SA, STV25235628SA, STV25235631SA, STV25184595SA, STV25235629SA, STV25235632SA, STV25235630SA, STV25276515SA, STV25225510SA, STV25276516SA, STV25225505SA, STV25225508SA, STV25225509SA, STV25225507SA, STV25225506SA, STV25113369SA, STV25204986SA, STV25205160SA, STV25205151SA, STV25205161SA, STV25062736SA, STV25256133SA, STV25225512SA, STV25296702SA, STV25296701SA,…
- Distribution
- Worldwide distribution.
- Recalling firm
- Wipro GE Healthcare Private Ltd., Bengaluru, India
- How the firm notified
- Letter
- FDA classified
- Oct 21, 2025
- GE Healthcare Carescape Central Station (CSCS) V3 MAS800 Desktop with software version V3.0.5; Part Number 5867474-04;…Z-0213-2026 · Class II · Sep 5, 2025 · Ongoing
Newest first, one product per recall event; search them all
- GE Giraffe Warmer, Model/Catalog Number M1118179-01090979. Infant radiant warmers provide infrared heat in a controlled…Z-3080-2024 · Class II · Aug 9, 2024 · Ongoing
- GE Healthcare neonatal incubator, models: a) Lullaby Incubator, b) Lullaby Incubator XP, c) Lullaby Incubator TRZ-0958-2024 · Class I · Dec 26, 2023 · Ongoing
FDA's weekly Enforcement Report, as openFDA publishes it. A recall is classified by how likely the product is to cause harm: Class I is the most serious. openFDA device enforcement reports