Arca
Skip to content
Tools
Tools

GE Healthcare neonatal incubator

Recall Z-0958-2024, part of FDA recall event 93780. See Wipro GE Healthcare's company page.

Wipro GE Healthcare Private Ltd. started this Class I device recall on Dec 26, 2023 (voluntary: firm initiated); FDA lists it as ongoing.

  • Classification: Class I
  • Started: Dec 26, 2023
  • Status: Ongoing
  • Products in this event: 6

Why It Was Recalled

Class I

GE HealthCare has become aware that the bedside panel of Care Plus, Care Plus¿ Model 1000, 2000, 3000 and 4000 Incubator, Lullaby Incubator, Lullaby TR Incubator, and Lullaby XP Incubator can be upright and look closed but not be latched if the red indicator is visible. The portholes also can look closed when not latched when the portholes are obstructed e.g., hood covers. If a patient impacts the bedside panel or porthole that is not latched, the panel can disengage and fall open, no longer protecting the patient from falling.

Record

Product
GE Healthcare neonatal incubator, models: a) Lullaby Incubator, b) Lullaby Incubator XP, c) Lullaby Incubator TR
Quantity
1410 units
Lots and codes
a) Lullaby Incubator, Model Numbers: M1151042, QCR031; b) Lullaby Incubator XP, Model Numbers: 8320, M1151135, QCR034; c) Lullaby Incubator TR, Model Numbers: M1151357 ALL SERIAL NUMBERS
Distribution
Worldwide distribution.
Recalling firm
Wipro GE Healthcare Private Ltd., Bengaluru, N/A, India
How the firm notified
Letter
FDA classified
Feb 9, 2024

Other Products in This Recall

Recall event 93780

Wipro GE Healthcare Private Ltd.'s Other Recalls

Newest first, one product per recall event; search them all

FDA's weekly Enforcement Report, as openFDA publishes it. A recall is classified by how likely the product is to cause harm: Class I is the most serious. openFDA device enforcement reports

Esc

Type a name: a tool, a company, a drug or brand, a condition, a gene or a state. Identifiers like K231234, NCT05012345 or NDA 215866 open their record.