GE Healthcare neonatal incubator
Recall Z-0958-2024, part of FDA recall event 93780. See Wipro GE Healthcare's company page.
Wipro GE Healthcare Private Ltd. started this Class I device recall on Dec 26, 2023 (voluntary: firm initiated); FDA lists it as ongoing.
- Classification: Class I
- Started: Dec 26, 2023
- Status: Ongoing
- Products in this event: 6
GE HealthCare has become aware that the bedside panel of Care Plus, Care Plus¿ Model 1000, 2000, 3000 and 4000 Incubator, Lullaby Incubator, Lullaby TR Incubator, and Lullaby XP Incubator can be upright and look closed but not be latched if the red indicator is visible. The portholes also can look closed when not latched when the portholes are obstructed e.g., hood covers. If a patient impacts the bedside panel or porthole that is not latched, the panel can disengage and fall open, no longer protecting the patient from falling.
- Product
- GE Healthcare neonatal incubator, models: a) Lullaby Incubator, b) Lullaby Incubator XP, c) Lullaby Incubator TR
- Quantity
- 1410 units
- Lots and codes
- a) Lullaby Incubator, Model Numbers: M1151042, QCR031; b) Lullaby Incubator XP, Model Numbers: 8320, M1151135, QCR034; c) Lullaby Incubator TR, Model Numbers: M1151357 ALL SERIAL NUMBERS
- Distribution
- Worldwide distribution.
- Recalling firm
- Wipro GE Healthcare Private Ltd., Bengaluru, N/A, India
- How the firm notified
- Letter
- FDA classified
- Feb 9, 2024
- GE Healthcare neonatal incubator, models: a) Care Plus, b) Care Plus 1000, c) Care Plus 2000, d) Care Plus…Z-0957-2024 · Class I · Dec 26, 2023 · Ongoing
- GE Healthcare Care Plus Hood Door FRU kit, Part Numbers: 6600-0082-850, 6600-0504-800, 6600-0148-800, and 6600-0970-500Z-0960-2024 · Class I · Dec 26, 2023 · Ongoing
- GE Healthcare Lullaby Incubator Model Hood Door FRU kit, Part Numbers: M1169767, M1160674, M1187512, M1206646, M1166822Z-0962-2024 · Class I · Dec 26, 2023 · Ongoing
- GE Healthcare Care Plus Hood Field Replacement Unit (FRU) kit, Part Numbers: 6600-0047-851, 6600-0033-850,…Z-0959-2024 · Class I · Dec 26, 2023 · Ongoing
- GE Healthcare Lullaby Incubator Model Hood FRU kit, Part Numbers: M1169566, M1160659, M1206259, M1206260, M1166836,…Z-0961-2024 · Class I · Dec 26, 2023 · Ongoing
- GE Healthcare Carescape Central Station (CSCS) V3 MAS800 Desktop with software version V3.0.5; Part Number 5867474-04;…Z-0213-2026 · Class II · Sep 5, 2025 · Ongoing
- GE Giraffe Warmer, Model/Catalog Number M1118179-01090979. Infant radiant warmers provide infrared heat in a controlled…Z-3080-2024 · Class II · Aug 9, 2024 · Ongoing
FDA's weekly Enforcement Report, as openFDA publishes it. A recall is classified by how likely the product is to cause harm: Class I is the most serious. openFDA device enforcement reports