GE Giraffe Warmer, Model/Catalog Number M1118179-01090979. Infant radiant warmers provide…
Recall Z-3080-2024, part of FDA recall event 95216. See Wipro GE Healthcare's company page.
Wipro GE Healthcare Private Ltd. started this Class II device recall on Aug 9, 2024 (voluntary: firm initiated); FDA lists it as ongoing.
- Classification: Class II
- Started: Aug 9, 2024
- Status: Ongoing
- Products in this event: 2
GE HealthCare has become aware that complete electrical safety testing was not conducted during manufacturing of certain Giraffe and Panda iRes Warmers. There are multiple electrical safety protections included in the product design. However, because the electrical safety testing in manufacturing was incomplete for these devices, there is the potential for leakage current to exceed IEC 60601 limits.
- Product
- GE Giraffe Warmer, Model/Catalog Number M1118179-01090979. Infant radiant warmers provide infrared heat in a controlled manner to neonates who are unable to thermo-regulate based on their own physiology. Infant radiant warmers may be used to facilitate the neonate's transition to the external environment or to provide a controlled open environment.
- Quantity
- 1
- Lots and codes
- Serial Number GBW24230381SA
- Distribution
- US and Australia, Denmark, France, Germany, Ireland, ITALY, New Zealand, Poland, Portugal, Slovenia, Spain, SWITZERLAND, United Arab Emirates, United Kingdom.
- Recalling firm
- Wipro GE Healthcare Private Ltd., Bengaluru, India
- How the firm notified
- Letter
- FDA classified
- Sep 20, 2024
- GE Panda iRES Warmer, Model/Catalog Numbers: 1) M1112198-01045518; 2) M1112198-01045579; 3) M1112198-01047148; …Z-3081-2024 · Class II · Aug 9, 2024 · Ongoing
- GE Healthcare Carescape Central Station (CSCS) V3 MAS800 Desktop with software version V3.0.5; Part Number 5867474-04;…Z-0213-2026 · Class II · Sep 5, 2025 · Ongoing
- GE Healthcare neonatal incubator, models: a) Lullaby Incubator, b) Lullaby Incubator XP, c) Lullaby Incubator TRZ-0958-2024 · Class I · Dec 26, 2023 · Ongoing