Arca
Skip to content
Tools
Tools

ABL700 Series Analyzer With Autocheck Module

Electrode Measurement, Blood-Gases (Pco2, Po2) And Blood Ph. 510(k) K992859, from Radiometer America, Inc. (Westlake, OH). Product code CHL, class 2.

As of Nov 18, 2026, K992859 is an FDA 510(k) held by Radiometer America, Inc., substantially equivalent on Jan 10, 2000 after 138 days, with 4 recalls tied to it in FDA's recall database.

  • Review time: 138 days
  • Recalls: 4
  • Cited as a predicate by: 2 devices

Predicate devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a predicate

2 later 510(k) summaries name this device as a predicate
  • AUTOCHECK6+K051928 · Radiometer Medical Aps · Oct 12, 2005
  • ABL800 FlexK041874 · Radiometer Medical Aps · Oct 8, 2004

Recalls

Recalls FDA ties to K992859, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
Jul 5, 2012Z-2163-2012, Z-2162-2012A customer has reported an error on an ABL8xx configured to a Hospital Information System using the data transfer protocol HL7 version 2.5. If the instrument transmits results to a HIS system and the connection is…Terminated
Sep 29, 2005Z-1386-06Under certain circumstances, the ABL 700 & 800 Series Blood Gas Analyzers may experience leakage current into the measuring system. Consequently, the analyzers intermittently provide incorrect (too low) result values…Terminated
Jun 17, 2005Z-1317-06Software defect. pO2 and pCO2 sample test results run on the firm's ABL700/800 Series Blood Gas Analyzers are not being properly flagged during the sample calibration phase.Terminated

Other CHL clearances

The newest 510(k)s for the same product code; their median review was 267 days; see every CHL device

Record

Type
Special
Received
Aug 25, 1999
Review panel
Clinical Chemistry
Regulation
21 CFR 862.1120

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files

Logos are each company's own site icon, matched by name through Wikidata.Company names and logos identify the companies in public FDA and registry records and do not imply endorsement.