ABL700 Series Blood Gas , Co-Oximetry, Electrolyte & Metabolite System

Electrode Measurement, Blood-Gases (Pco2, Po2) And Blood Ph. 510(k) K980130, from Radiometer America, Inc. (Westlake, OH). Product code CHL, class 2.

As of Nov 18, 2026, K980130 is an FDA 510(k) held by Radiometer America, Inc., substantially equivalent on Feb 9, 1998 after 26 days, with 5 recalls tied to it in FDA's recall database.

  • Review time: 26 days
  • Recalls: 5
  • Cited as a predicate by: 3 devices

Predicate devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a predicate

3 later 510(k) summaries name this device as a predicate

Recalls

Recalls FDA ties to K980130, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
Jul 5, 2013Z-1075-2014RADIOMETER has become aware that some D788 pCO2 membranes can cause biased measurement results on patient results as well as QC results.Terminated
Apr 7, 2011Z-2571-2011Based on internal testing and user/customer complaint reports received, the firm determined that wrinkles may arise on the outer membrane of the calcium membrane units upon installation. Retention of fluids between the…Terminated
Sep 29, 2005Z-1386-06Under certain circumstances, the ABL 700 & 800 Series Blood Gas Analyzers may experience leakage current into the measuring system. Consequently, the analyzers intermittently provide incorrect (too low) result values…Terminated
Sep 19, 2005Z-1323-06Reference electrodes for the firm's ABL 700/800 Blood Gas Analyzers are oversized and will not fit into the instrument.Terminated
Feb 1, 2005Z-1314-06The CI membrane units generate too low values for status and sensitivity during calibration.Terminated

Other CHL clearances

The newest 510(k)s for the same product code; their median review was 267 days; see every CHL device

Record

Type
Traditional
Received
Jan 14, 1998
Review panel
Clinical Chemistry
Regulation
21 CFR 862.1120

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files