Sterrad 100S Sterilizer

Sterilizer, Chemical. 510(k) K991999, from Advanced Sterilization Products (Irvin, CA). Product code MLR, class 2.

As of Nov 25, 2026, K991999 is an FDA 510(k) held by Advanced Sterilization Products, substantially equivalent on Jun 15, 2000 after 367 days, with 6 recalls tied to it in FDA's recall database.

  • Review time: 367 days
  • Recalls: 6
  • Cited as a predicate by: 2 devices

Indications for use

As the 510(k) summary states them, read from FDA's summary PDF

The STERRAD ® 100S Sterilizer is designed for sterilization of both metal and nonmetal medical devices at low temperatures. Because the cycle operates within a dry environment and at low temperatures, it is especially suitable for instruments sensitive to heat and moisture.

Cited as a predicate

2 later 510(k) summaries name this device as a predicate

Recalls

Recalls FDA ties to K991999, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
Jun 19, 2012Z-2337-2012Advanced Sterilization Products (ASP) is recalling (removing) the Sterrad 100S and Sterrad NX System cassettes because the camera used to verify printing on the Sterrad System cassettes was inadvertently disabled. The…Terminated
Oct 17, 2008Z-0340-20091) Inability of the sterilizer to detect when an injection takes place without hydrogen peroxide being transferred to the vaporizer bowl. This situation can result in a cycle completion with the injection of…Terminated
Feb 25, 2008Z-1628-2008Oil mist filter fails allowing oil mist to be emitted into the vicinity of the sterilizer, which may result in the release of a mist, "haze," or "smoke."Terminated
Nov 12, 2007Z-0845-2008User Guide Update: based on user reports about contact with residual hydrogen peroxide from instruments, pouches and trays after sterilization and subsequent light colored residue on these devices after sterilization…Terminated
Nov 12, 2007Z-0844-2008Outdated Compatibility Lists: Recalling firm has decided to discontinue dissemination of Compatible Medical Device Reference Lists and instrument assessment activitiesTerminated
Sep 16, 2005Z-0197-06The STERRAD 100 and the STERRAD 100S vaporizer plates are prone to displacement allowing micro-droplets of H2O2 to contact the load. If the vaporizer plate is dislodged, there is an increased chance for operators to…Terminated

Other MLR clearances

The newest 510(k)s for the same product code; their median review was 216 days; see every MLR device

Record

Type
Traditional
Received
Jun 14, 1999
Review panel
General Hospital
Regulation
21 CFR 880.6860

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files