Arca
Skip to content
Tools
Tools

Gammamed Plus High Dose Rate Remote Afterloading System

System, Applicator, Radionuclide, Remote-Controlled. 510(k) K983436, from Gammamed USA (Ann Arbor, MI). Product code JAQ, class 2.

As of Nov 20, 2026, K983436 is an FDA 510(k) held by Gammamed USA, substantially equivalent on Oct 29, 1998 after 30 days, with 17 recalls tied to it in FDA's recall database.

  • Review time: 30 days
  • Recalls: 17
  • Cited as a predicate by: 13 devices

Indications for use

As the 510(k) summary states them, read from FDA's summary PDF

The intended use of the GammaMed Plus transportable high-dose-rate remotely controlled afterloading brachytherapy device is for the treatment of cancer by intracavitary, interstitial, intralumenal and intraoperative irradiation.

Predicate devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a predicate

13 later 510(k) summaries name this device as a predicate; the 12 newest

Recalls

Recalls FDA ties to K983436, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
Sep 28, 2012Z-0607-2013The guide tube which is part of the Segmented Cervix Applicator Set is difficult to disconnect from cervical sleeve.Terminated
Jul 13, 2011Z-2995-2011, Z-2998-2011, Z-3000-2011, Z-2999-2011, Z-2996-2011, Z-2997-2011Potential of weld failure of the Titanium intrauterine probe at the cervical stopper.Terminated
Jul 13, 2011Z-2992-2011, Z-2991-2011, Z-2993-2011Needle tip may separate from the needle shaft such that the tip may detach in the patient upon withdrawal.Terminated
Jul 13, 2011Z-3013-2011Certain GammaMed Flexible applicator probes may not have been heat annealed during manufacturing to prevent shrinkage after autoclave.Terminated
Sep 30, 2008Z-0328-2012, Z-0327-2012, Z-0326-2012Plastic needle tip could possibly crack or separate.Terminated
Aug 5, 2008Z-0295-2012GammaMed Flexible Applicator Probes may not have been heat annealed during manufacturing to prevent shrinkage after autoclave.Terminated
May 1, 2006Z-1268-06Medical device software for brachytherapy may cause erroneous data to be recorded and affect patient radiation treatments. Erroneous data may be listed on the treatment history report, and the default step size may lead…Terminated
Apr 14, 2005Z-0921-05A defective treatment applicator in a cancer therapy medical device may become separated potentially causing patient injury.Terminated

Other JAQ clearances

The newest 510(k)s for the same product code; their median review was 98 days; see every JAQ device

Record

Type
Special
Received
Sep 29, 1998
Review panel
Radiology
Regulation
21 CFR 892.5700

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files