Arca
Skip to content
Tools
Tools

Varian Intracavitary Brachytherapy Applicators

System, Applicator, Radionuclide, Remote-Controlled. 510(k) K033371, from Varian Medical Systems (Palo Alto, CA). Product code JAQ, class 2.

As of Nov 20, 2026, K033371 is an FDA 510(k) held by Varian Medical Systems, substantially equivalent on Dec 24, 2003 after 64 days, with 2 recalls tied to it in FDA's recall database.

  • Review time: 64 days
  • Recalls: 2
  • Cited as a predicate by: 8 devices

Indications for use

As the 510(k) summary states them, read from FDA's summary PDF

The applicators will be used with the VariSource High Dose Rate Afterloaders to deliver brachytherapy treatment for gynecological and rectal applications. The applicators will be used in medical intracavitary for treatment of cancerous tumors.

Predicate devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a predicate

8 later 510(k) summaries name this device as a predicate

Recalls

Recalls FDA ties to K033371, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
Jul 13, 2011Z-3055-2011Blocking washers on the Flexible Probe with Blocking washers may shift and cause potential for treatment length error.Terminated
May 20, 2010Z-0282-2012The Instructions For Use ( IFU) distributed with product prior to 2004 lacks the correct limit to the sterilization cycles and checks regarding the firm seating of the blocking washer.Terminated

Other JAQ clearances

The newest 510(k)s for the same product code; their median review was 98 days; see every JAQ device

Record

Type
Traditional
Received
Oct 21, 2003
Review panel
Radiology
Regulation
21 CFR 892.5700

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files