Sterrad 50 Sterilizer
Sterilizer, Chemical. 510(k) K981625, from Advanced Sterilization Products (Irvin, CA). Product code MLR, class 2.
As of Nov 25, 2026, K981625 is an FDA 510(k) held by Advanced Sterilization Products, substantially equivalent on Jan 11, 1999 after 249 days, with 5 recalls tied to it in FDA's recall database.
- Review time: 249 days
- Recalls: 5
- Cited as a predicate by: 2 devices
The STERRAD® 50 Sterilizer is designed for sterilization of both metal and nonmetal medical devices at low temperatures. Because the cycle operates within a dry environment and at low temperatures, it is especially suitable for instruments sensitive to heat and moisture.
- Sterrad Bi Test PackK993775 · Advanced Sterilization Products · Jun 1, 2000
- Sterrad 50 & Sterrad 100S SterilizersK023290 · Advanced Sterilization Products · Apr 3, 2003
| Mar 9, 2009 | Z-1419-2009ASP has identified a secondary, lower-occurring cause of oil mist within certain STERRAD NX Systems and STERRAD 50 Systems due to potentially defective oil fill plugs. | Terminated |
| Jan 12, 2009 | Z-1570-2009The capacitor in the vacuum pump component within the Sterrad 50 System could overheat and potentially create a hazardous condition inside the system, resulting in smoke and/or fire. | Terminated |
| Feb 25, 2008 | Z-1629-2008Oil mist filter fails allowing oil mist to be emitted into the vicinity of the sterilizer, which may result in the release of a mist, "haze," or "smoke." | Terminated |
| Nov 12, 2007 | Z-0845-2008User Guide Update: based on user reports about contact with residual hydrogen peroxide from instruments, pouches and trays after sterilization and subsequent light colored residue on these devices after sterilization… | Terminated |
| Nov 12, 2007 | Z-0844-2008Outdated Compatibility Lists: Recalling firm has decided to discontinue dissemination of Compatible Medical Device Reference Lists and instrument assessment activities | Terminated |
- V-Pro maX 2 Low Temperature Sterilization SystemK263530 · Steris · Oct 29, 2026
- STERRAD 100NX Sterilization System with ALLClear Technology (10104)K252843 · Advanced Sterilization Products, Inc. · Apr 15, 2026
- STERRAD® 100NX Sterilizer with ALLClear™ Technology (10104)K250802 · Advanced Sterilization Products, Inc. · Apr 14, 2025
- LOWTEM Crystal 120K231893 · Lowtem Co., Ltd. · Sep 18, 2024
- STERRAD® 100NX Sterilizer with ALLClear™ Technology and ULTRA GI™ Cycle (10104); APTIMAX™ Instrument Tray for ULTRA GI™…K234082 · Advanced Sterilization Products · Jul 5, 2024
- SteroScope® Sterilization Technology SystemK233762 · Ideate Medical, Inc. · Jun 27, 2024
- V-PRO maX 2 Low Temperature Sterilization SystemK233065 · Steris · Oct 20, 2023
- V-PRO maX 2 Low Temperature Sterilization SystemK223476 · STERIS Corporation · Aug 7, 2023
- Document
- 510(k) summary (PDF) ↗
- Type
- Traditional
- Received
- May 7, 1998
- Review panel
- General Hospital
- Regulation
- 21 CFR 880.6860
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files