Orthofix Modulsystem
Pin, Fixation, Threaded. 510(k) K955848, from Orthofix, Inc. (Rockville, MD). Product code JDW, class 2.
As of Nov 20, 2026, K955848 is an FDA 510(k) held by Orthofix, Inc., substantially equivalent on Mar 20, 1996 after 84 days, with 23 recalls tied to it in FDA's recall database.
- Review time: 84 days
- Recalls: 23
- Cited as a predicate by: 11 devices
- XT3 SystemK241357 · Biodynamik, Inc. · Jan 22, 2025
- Fitbone TrochantericK233867 · Orthofix Srl · Jun 18, 2024
- Pitkar External Fixation System - Rail MechanismK200518 · S.H.Pitkar Orthotools Pvt. , Ltd. · Jul 29, 2020
- PRECICE Plating SystemK192181 · Nuvasive Specialized Orthopedics, Inc. · Nov 7, 2019
- Orthofix Tl-Hex True Lok Hexalobe System (Tl-Hex)K141078 · Orthofix Srl · Sep 2, 2014
- Newfix External Fixation SystemK122210 · Tecnologia Y Diseno Industrial, S.A DE C.V · Apr 19, 2013
- Orthofix Galaxy Fixation SystemK113770 · Orthofix Srl · Sep 4, 2012
- Newfix External Fixation SystemK101338 · Technologia Y Diseno Industrial, S.A DE C.V · Mar 14, 2011
- Biomet Vision Footring SystemK071395 · Biomet Manufacturing Corp · Jul 9, 2007
- Reprocessed External Fixation DeviceK023714 · Alliance Medical Corp. · Feb 3, 2003
- Iskd (Intramedullary Skeletal Kinetic Distractor)K010322 · Orthodyne, Inc. · May 2, 2001
| Sep 27, 2018 | Z-0393-2020, Z-0380-2020, Z-0382-2020, Z-0391-2020, Z-0392-2020, Z-0389-2020, Z-0387-2020, Z-0386-2020, Z-0374-2020, Z-0377-2020, Z-0378-2020, Z-0384-2020, Z-0388-2020, Z-0376-2020, Z-0373-2020, Z-0375-2020, Z-0383-2020, Z-0379-2020, Z-0381-2020, Z-0385-2020, Z-0390-2020As a result of complaint investigations it was determined that batches of drill bits have a geometrical detail that does not correspond to the intended design. This could result in lower cutting performance. | Terminated |
| Feb 22, 2010 | Z-1435-2010, Z-1434-2010During standard quality controls, firm identified a potential for breakage of the XCaliber clamp when locking the cam. | Terminated |
- XT3 System; XT3 System, Mini; XT3 System, 40mm; XT3 System, Mini, 40mmK260073 · Biodynamik, Inc. · Feb 10, 2026
- Meduloc Intramedullary Fracture Fixation (IFF) SystemK250316 · Meduloc, LLC · Oct 29, 2025
- XT3 SystemK241357 · Biodynamik, Inc. · Jan 22, 2025
- MAVERICK Mini External Fixation SystemK223112 · Smith & Nephew, Inc. · Dec 8, 2022
- MAVERICK External Fixation SystemK213874 · Smith & Nephew, Inc. · Feb 11, 2022
- Steinmann Pins and Kirschner WiresK211508 · Stryker GmbH · Oct 5, 2021
- Extended HA Half PinsK210784 · Smith & Nephew, Inc. · Apr 2, 2021
- Document
- 510(k) summary (PDF) ↗
- Type
- Traditional
- Received
- Dec 27, 1995
- Review panel
- Orthopedic
- Regulation
- 21 CFR 888.3040
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files