Arca
Skip to content

XT3 System

Pin, Fixation, Threaded. 510(k) K241357, from Biodynamik, Inc. (Lake Forest, CA). Product code JDW, class 2.

As of Nov 16, 2026, K241357 is an FDA 510(k) held by Biodynamik, Inc., substantially equivalent on Jan 22, 2025 after 253 days, with no recalls tied to it in FDA's recall database.

  • Substantially equivalent: Jan 22, 2025
  • Review time: 253 days
  • Recalls: 0
  • Cited as a predicate by: 1 device

Indications for Use

As the 510(k) summary states them, read from FDA's summary PDF

The BioDynamik XT3 System is indicated for fracture fixation and bone transport of the tibia in adults.

Predicate Devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a Predicate

1 later 510(k) summary names this device as a predicate

Other JDW Clearances

The newest 510(k)s for the same product code; their median review was 66 days; see every JDW device

Record

Type
Traditional
Received
May 14, 2024
Review panel
Orthopedic
Regulation
21 CFR 888.3040

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files

Logos are each company's own site icon, matched by name through Wikidata.Company names and logos identify the companies in public FDA and registry records and do not imply endorsement.