PRECICE Plating System
Appliance, Fixation, Nail/Blade/Plate Combination, Multiple Component. 510(k) K192181, from Nuvasive Specialized Orthopedics, Inc. (Alisa Viejo, CA). Product code KTT, class 2.
As of Nov 17, 2026, K192181 is an FDA 510(k) held by Nuvasive Specialized Orthopedics, Inc., substantially equivalent on Nov 7, 2019 after 87 days, with 1 recall tied to it in FDA's recall database.
- Substantially equivalent: Nov 7, 2019
- Review time: 87 days
- Recalls: 1
- Cited as a predicate by: 2 devices
Indications for Use
As the 510(k) summary states them, read from FDA's summary PDFThe PRECICE ® Plate System is indicated for limb lengthening, open and closed fracture fixation, pseudoarthrosis, mal- unio ns and non- unions of long bones in pediatrics and small stature adult patients.
Predicate Devices
The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF- Inter-Os Limb LengthenerK031875 · Inter-Os Technologies · Oct 22, 2003
- Orthopediatrics Fracture And Osteotomy Bone Plate SystemK111086 · OrthoPediatrics Corp. · Jun 17, 2011
- PRECICE STRYDE SystemK180503 · Nuvasive Specialized Orthopedics, Inc. · Apr 3, 2018
- Orthofix ModulsystemK955848 · Orthofix, Inc. · Mar 20, 1996
Cited as a Predicate
2 later 510(k) summaries name this device as a predicate- External Remote Controller ERC 4PK202348 · Nuvasive Specialized Orthopedics, Inc. · Dec 22, 2020
- Precice® ScrewsK193617 · Nuvasive Specialized Orthopedics, Inc. · May 8, 2020
Recalls
Recalls FDA ties to K192181, newest first; each opens FDA's recall record| Started | Reason | Status |
|---|---|---|
| Feb 12, 2021 | Z-1325-2021Due to complaints of adverse events potentially related to biological safety. The complaints included report of pain and bony abnormalities at the interface between the telescoping nail segments. | Open, Classified |
Other KTT Clearances
The newest 510(k)s for the same product code; their median review was 84 days; see every KTT device- Response Ortho Smart Fixator - Hexapod SystemK263083 · Response Ortho Solutions, LLC · Oct 2, 2026
- MAXFRAME AUTOSTRUT with Bluetooth Multi-Axial Correction SystemK254087 · Synthes (USA) Products, LLC · Jun 11, 2026
- DYNEX® External Fixation SystemsK260900 · Vilex, LLC · Apr 17, 2026
- QuikFix External Fixator: Knee-Spanning PackK260112 · Stabiliz Orthopaedics, Inc. · Mar 23, 2026
- Excelsior SystemK253291 · Blue Ocean Global · Jan 20, 2026
- Hoffmann LRF SystemK253202 · Stryker GmbH · Dec 19, 2025
- Monkey Rings External Ring Fixation SystemK253613 · Paragon 28, Inc. · Dec 19, 2025
- Medusa Orthopedics Boa External Fixation SystemK252555 · Medusa Orthopedics, LLC · Nov 25, 2025
Record
- Document
- 510(k) summary (PDF) ↗
- Type
- Traditional
- Received
- Aug 12, 2019
- Review panel
- Orthopedic
- Regulation
- 21 CFR 888.3030
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files