Hydradjust Iv Urological Table
Table, Radiographic, Tilting. 510(k) K943581, from Mallinckrodt Group, Inc. (Cincinnati, OH). Product code IXR, class 2.
As of Nov 17, 2026, K943581 is an FDA 510(k) held by Mallinckrodt Group, Inc., substantially equivalent on Jan 4, 1995 after 166 days, with 3 recalls tied to it in FDA's recall database.
- Substantially equivalent: Jan 4, 1995
- Review time: 166 days
- Recalls: 3
- Cited as a predicate by: 3 devices
Cited as a Predicate
3 later 510(k) summaries name this device as a predicate- Uroview 2800K993687 · Ge Dec Medical Systems · Dec 21, 1999
- Liebel-Flarsheim Direct Digital Imaging SystemK121838 · Liebel-Flarsheim · Jul 11, 2012
- UrologicK011311 · Pck Electronic Industry and Trade Co, Ltd. · Jul 10, 2001
Recalls
Recalls FDA ties to K943581, newest first; each opens FDA's recall record| Feb 27, 2017 | Z-2833-2018Software issue. The difference between the display and dosimeter readings In the Child/Pediatric automatic exposure mode is in the range of 52-65%. The display in all other modes deviated by less than 35% from the… | Terminated |
| Apr 25, 2013 | Z-1753-2013A complaint was received in which a hospital physicist indicated the dosage was too high on the HUT system. During the service call the systems Maximum Fluoro Dose was found to be 30 R/min which is three times the… | Terminated |
| Jun 14, 2010 | Z-1904-2010Covidien received complaints of the Sedecal Generator failing and producing smoke at customer sites. The failure modes are located in different parts of the generator and will lead to the generator becoming inoperative. | Terminated |
Record
- Type
- Traditional
- Received
- Jul 22, 1994
- Review panel
- Radiology
- Regulation
- 21 CFR 892.1980
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files