Arca
Skip to content

Liebel-Flarsheim Direct Digital Imaging System

Solid State X-Ray Imager (Flat Panel/Digital Imager). 510(k) K121838, from Liebel-Flarsheim (Cincinnati, OH). Product code MQB, class 2.

As of Nov 17, 2026, K121838 is an FDA 510(k) held by Liebel-Flarsheim, substantially equivalent on Jul 11, 2012 after 19 days, with 2 recalls tied to it in FDA's recall database.

  • Substantially equivalent: Jul 11, 2012
  • Review time: 19 days
  • Recalls: 2

Indications for Use

As the 510(k) summary states them, read from FDA's summary PDF

The Liebel-Flarsheim Direct Digital Imaging System is a solid state detector X-ray system that facilitates digital radiologic and/or fluoroscopic procedures requiring a beam of diagnostic quality radiation and a flat imaging table primarily for urological applications such as functional x-ray diagnostics, endourology and minimal invasive urology/surgery.

Predicate Devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Recalls

Recalls FDA ties to K121838, newest first; each opens FDA's recall record
Mar 11, 2021Z-1992-2021Failure to comply with the performance standard for fluoroscopy systems. Per 21 CFR 1020.32(k)(6), the displayed AKR and cumulative air kerma shall not deviate from the actual values by more than 35 percent. The dosage…Terminated
Feb 27, 2017Z-2834-2018Software issue. The difference between the display and dosimeter readings In the Child/Pediatric automatic exposure mode is in the range of 52-65%. The display in all other modes deviated by less than 35% from the…Terminated

Other MQB Clearances

The newest 510(k)s for the same product code; their median review was 179 days; see every MQB device

Record

Type
Traditional, third-party review
Received
Jun 22, 2012
Review panel
Radiology
Regulation
21 CFR 892.1680

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files