Liebel-Flarsheim Direct Digital Imaging System
Solid State X-Ray Imager (Flat Panel/Digital Imager). 510(k) K121838, from Liebel-Flarsheim (Cincinnati, OH). Product code MQB, class 2.
As of Nov 17, 2026, K121838 is an FDA 510(k) held by Liebel-Flarsheim, substantially equivalent on Jul 11, 2012 after 19 days, with 2 recalls tied to it in FDA's recall database.
- Substantially equivalent: Jul 11, 2012
- Review time: 19 days
- Recalls: 2
Indications for Use
As the 510(k) summary states them, read from FDA's summary PDFThe Liebel-Flarsheim Direct Digital Imaging System is a solid state detector X-ray system that facilitates digital radiologic and/or fluoroscopic procedures requiring a beam of diagnostic quality radiation and a flat imaging table primarily for urological applications such as functional x-ray diagnostics, endourology and minimal invasive urology/surgery.
Predicate Devices
The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF- Hydradjust Iv Urological TableK943581 · Mallinckrodt Group, Inc. · Jan 4, 1995
- Uroskop OmniaK101491 · Siemens Medical Solutions USA, Inc. · Jun 30, 2010
- Nexus Drf Digital X-Ray Imaging SystemK103416 · Infimed, Inc. · Mar 17, 2011
Recalls
Recalls FDA ties to K121838, newest first; each opens FDA's recall record| Mar 11, 2021 | Z-1992-2021Failure to comply with the performance standard for fluoroscopy systems. Per 21 CFR 1020.32(k)(6), the displayed AKR and cumulative air kerma shall not deviate from the actual values by more than 35 percent. The dosage… | Terminated |
| Feb 27, 2017 | Z-2834-2018Software issue. The difference between the display and dosimeter readings In the Child/Pediatric automatic exposure mode is in the range of 52-65%. The display in all other modes deviated by less than 35% from the… | Terminated |
Other MQB Clearances
The newest 510(k)s for the same product code; their median review was 179 days; see every MQB device- 0909FCE, 0909FCE-HSK253800 · Rayence Co., Ltd. · Jul 31, 2026
- Vivix-S 4386W (FXRD-4386WA)K260820 · Vieworks Co., Ltd. · Jun 18, 2026
- Universal 1417PIK252996 · Konica Minolta Healthcare Americas, Inc. · Jun 12, 2026
- Lux HD 2530 detector (Lux HD 2530)K252911 · Iray Imaging Technology (Haining) Limited · Oct 9, 2025
- Yushan X-Ray Flat Panel DetectorK250211 · Innocare Optoelectronics Corp. · Jul 22, 2025
- Skr 3000K250665 · Konica Minolta, Inc. · Jun 17, 2025
- Wireless/ Wired X-Ray Flat Panel DetectorsK243734 · Allengers Medical Systems Limited · Apr 18, 2025
- Expd 114; Expd 114G; Expd 114P; Expd 114PGK242770 · DRTECH Corporation · Mar 20, 2025
Record
- Document
- 510(k) summary (PDF) ↗
- Type
- Traditional, third-party review
- Received
- Jun 22, 2012
- Review panel
- Radiology
- Regulation
- 21 CFR 892.1680
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files