Uroview 2800
Interventional Fluoroscopic X-Ray System. 510(k) K993687, from Ge Dec Medical Systems (Salt Lake City, UT). Product code OWB, class 2.
As of Nov 17, 2026, K993687 is an FDA 510(k) held by Ge Dec Medical Systems, substantially equivalent on Dec 21, 1999 after 50 days, with 13 recalls tied to it in FDA's recall database.
- Substantially equivalent: Dec 21, 1999
- Review time: 50 days
- Recalls: 13
Indications for Use
As the 510(k) summary states them, read from FDA's summary PDFThe UroView 2800 is designed to provide fluoroscopic and spot-film imaging of the patient during diagnostic, surgical, and interventional procedures. Clinical applications include but are not limited to urologic and endoscopic procedures. The system may be used for other imaging applications at the physician's discretion.
Predicate Devices
The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF- Uroview Model 2600K940295 · Ge Dec Medical Systems · Aug 22, 1994
- Hydradjust Iv Urological TableK943581 · Mallinckrodt Group, Inc. · Jan 4, 1995
- Hydra Vision Urological X-Ray SystemK904145 · Mallinckrodt Group, Inc. · Dec 13, 1990
Recalls
Recalls FDA ties to K993687, newest first; each opens FDA's recall record| Jul 22, 2016 | Z-2628-2016A firmware issue in the supplied LCD workstation monitor(s). Under some circumstances, the monitor(s) may appear dark with no video displayed. | Terminated |
| Jul 22, 2016 | Z-2629-2016A firmware issue in the supplied LCD workstation monitor(s). Under some circumstances, the monitor(s) may appear dark with no video displayed. | Terminated |
| Oct 3, 2014 | Z-0192-2015GE Healthcare initiated a field correction because prior revision of components were potentially installed on certain OEC 9800, FlexiView 8800, OEC MiniView 6800 and OEC UroView 2800 systems. | Terminated |
| Aug 29, 2008 | Z-1730-2009GE Healthcare has found that the sterile products manufactured by Contour Fabricators, Inc. (CFI) for GE healthcare (GEHC) may have sterile seals that could be compromised during the shelf life of the products. This… | Terminated |
| Aug 15, 2008 | Z-1706-2009GE Healthcare has found that the sterile products manufactured by Contour Fabricators, Inc. (CFI) for GE healthcare (GEHC) may have seals that could be compromised during the shelf life of the products. This issue could… | Terminated |
| Aug 15, 2008 | Z-1708-2009GE Healthcare has found that the sterile products manufactured by Contour Fabricators, Inc. (CFI) for GE healthcare (GEHC) may have seals that could be compromised during the shelf life of the products. This issue could… | Terminated |
| Aug 15, 2008 | Z-1707-2009GE Healthcare has found that the sterile products manufactured by Contour Fabricators, Inc. (CFI) for GE healthcare (GEHC) may have seals that could be compromised during the shelf life of the products. This issue could… | Terminated |
| Aug 15, 2008 | Z-1709-2009GE Healthcare has found that the sterile products manufactured by Contour Fabricators, Inc. (CFI) for GE healthcare (GEHC) may have seals that could be compromised during the shelf life of the products. This issue could… | Terminated |
| Jun 12, 2007 | Z-0852-2007X-ray systems were not configured to display Air Kerma Rate and Cumulative Air Kerma. | Terminated |
| Mar 27, 2007 | Z-2127-2012Please be aware that this is not a new recall. The firm has taken action; but, due to administrative issues this recall is now being classified by the Agency. GE OEC had discovered an issue with the hard disk drives… | Terminated |
| Jan 15, 2007 | Z-0514-2007Uroview 2800 system with dual monitors may pose a safety risk to patients if the monitors become disconnected and fall. | Terminated |
| Jan 15, 2007 | Z-0515-2007Uroview 2800 system with 3 Phase Power Distribution Box (Wonder Box) option can present a shock hazard to the user if the access door to the power box is not completely closed when the unit is powered on. . | Terminated |
| Nov 20, 2006 | Z-1865-2012Recalled in November 2006 due to several intermittent potential safety issues including navigation tracking inaccuracy, system Lok-up, incorrect image display, incorrect dose area calculated air kerma in "film mode" and… | Terminated |
Other OWB Clearances
The newest 510(k)s for the same product code; their median review was 231 days; see every OWB device- uAngio AVIVA CXK253726 · Shanghai United Imaging Healthcare Co., Ltd. · Aug 21, 2026
- Extron 3;EXTRON 6K261413 · DRTECH Corporation · Aug 19, 2026
- Zenition 30K254205 · Philips Medical Systems Nederland B.V. · Aug 7, 2026
- Shield System (Siemens 30x40)K260607 · Radiaction , Ltd. · Jul 31, 2026
- SmartCT (R3.0 (L10))K253670 · Philips Medical Systems B.V. · Jul 10, 2026
- C-beamK252735 · Pulmera, Inc. · May 22, 2026
- Intelligent NRK252503 · Canon, Inc. · Apr 30, 2026
- ARTIS icono floor; ARTIS icono biplane; ARTIS icono ceiling; ARTIS phenoK254173 · Siemens Medical Solutions USA, Inc. · Apr 24, 2026
Record
- Document
- 510(k) summary (PDF) ↗
- Type
- Abbreviated
- Received
- Nov 1, 1999
- Review panel
- Radiology
- Regulation
- 21 CFR 892.1650
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files