Arca
Skip to content
Tools
Tools

Baxter Personal Cycler Autommated Peritoneal Dial

System, Peritoneal, Automatic Delivery. 510(k) K923065, from Baxter Healthcare Corp (Round Lake, IL). Product code FKX, class 2.

As of Nov 27, 2026, K923065 is an FDA 510(k) held by Baxter Healthcare Corp, substantially equivalent on Mar 4, 1994 after 617 days, with 4 recalls tied to it in FDA's recall database.

  • Review time: 617 days
  • Recalls: 4
  • Cited as a predicate by: 6 devices

Cited as a predicate

6 later 510(k) summaries name this device as a predicate

Recalls

Recalls FDA ties to K923065, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
Jul 6, 2012Z-2132-2012Baxter discovered that a Baxter technical service representative was providing incorrect instructions for inspection of solution bags for a blocked frangible prior to connection to the Automated PD Set with Cassette.Terminated
Nov 20, 2007Z-0946-2008Unrecorded Infusion; If the system is powered down or a power failure occurs during a fill cycle, the HomeChoice system may not record the last 1-4 pump strokes infused into the patient when the power is restored. This…Terminated
Jun 21, 2005Z-1018-05The screws that secure the Power Entry Module (PEM) to the base of the HomeChoice device are not reliably grounded to the device's ground system, posing an electric shock hazard.Terminated
May 24, 2004Z-1012-04A software defect could result in a patient experiencing an overfill with software versions 8.5 and higher. The possibility of an overfill can occur when the Initial Drain Alarm Volume is set to OFF and a change to the…Terminated

Other FKX clearances

The newest 510(k)s for the same product code; their median review was 223 days; see every FKX device

Record

Type
Traditional
Received
Jun 25, 1992
Review panel
Gastroenterology, Urology
Regulation
21 CFR 876.5630

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files