Homechoice Personal Cycler Peritoneal Dialysis System, Models 5C8310, 5C8302, 5C4471, 5C4469

System, Peritoneal, Automatic Delivery. 510(k) K012988, from Baxter Healthcare Corp (Mcgaw Park, IL). Product code FKX, class 2.

As of Nov 27, 2026, K012988 is an FDA 510(k) held by Baxter Healthcare Corp, substantially equivalent on Dec 5, 2001 after 90 days, with 9 recalls tied to it in FDA's recall database.

  • Review time: 90 days
  • Recalls: 9
  • Cited as a predicate by: 5 devices

Indications for use

As the 510(k) summary states them, read from FDA's summary PDF

HomeChoice® Personal Cycler Peritoneal Dialysis System The HomeChoice® Personal Cycler Peritoneal Dialysis System is intended for automatic control of dialysate solution exchanges in the treatment of pediatric and adult renal failure patients undergoing peritoneal dialysis.

Predicate devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a predicate

5 later 510(k) summaries name this device as a predicate

Recalls

Recalls FDA ties to K012988, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
Nov 20, 2007Z-0947-2008Unrecorded Infusion; If the system is powered down or a power failure occurs during a fill cycle, the HomeChoice system may not record the last 1-4 pump strokes infused into the patient when the power is restored. This…Terminated
Sep 12, 2005Z-0139-06The Spanish language patient operating manual called the Patient At-Home Guide, was not updated when several warnings were added to the English version that may assist the operator in the safe and effective operation of…Terminated
Jun 21, 2005Z-1019-05The screws that secure the Power Entry Module (PEM) to the base of the HomeChoice device are not reliably grounded to the device's ground system, posing an elcetric shock hazard.Terminated
Dec 23, 2004Z-0397-05A software anomaly may allow concentrated solution to be infused into the patient when the HomeChoice unit is used with Physioneal Clear-Flex PD products.Terminated
Nov 1, 2004Z-0189-05, Z-0188-05The on/off power switch is not grounded as a separate component in the device, and can loosen from its housing, which may result in an electrical shock to the user.Terminated
May 24, 2004Z-1013-04A software defect could result in a patient experiencing an overfill with software versions 8.5 and higher. The possibility of an overfill can occur when the Initial Drain Alarm Volume is set to OFF and a change to the…Terminated
Dec 23, 2002Z-0565-03, Z-0564-03Potential for low ultrafiltrate volumes when using the Low Fill Mode during Continuous Cycling Peritoneal Dialysis (CCPD) TherapyTerminated

Other FKX clearances

The newest 510(k)s for the same product code; their median review was 223 days; see every FKX device

Record

Type
Traditional
Received
Sep 6, 2001
Review panel
Gastroenterology, Urology
Regulation
21 CFR 876.5630

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files