Homechoice/Homechoice Pro Automated Personal Cycler Peritoneal Dialysis System Model 5C4471, 5C8310, 5C4471R and 5C8310R

System, Peritoneal, Automatic Delivery. 510(k) K102936, from Baxter Healthcare Corp (Mcgaw Park, IL). Product code FKX, class 2.

As of Nov 27, 2026, K102936 is an FDA 510(k) held by Baxter Healthcare Corp, substantially equivalent on Mar 30, 2011 after 177 days, with 11 recalls tied to it in FDA's recall database.

  • Review time: 177 days
  • Recalls: 11
  • Cited as a predicate by: 2 devices

Predicate devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a predicate

2 later 510(k) summaries name this device as a predicate

Recalls

Recalls FDA ties to K102936, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
Oct 10, 2018Z-0030-2020Potential presence of leaks from the cassettes, which may be present due to an incomplete seal of the sheeting to the cassette. A leaking cassette could lead to contamination of the sterile fluid path with…Terminated
Jul 5, 2017Z-3040-2017, Z-3041-2017Baxter Healthcare Corporation has been made aware that users may not be following the instructions in the Operators Manual and incorrectly (using sharp objects) opening disposable set packaging while setting up their…Terminated
Dec 16, 2016Z-1056-2017Customer complaints received for the presence of leaksTerminated
Jun 29, 2015Z-0291-2016Loud operating sounds, which was unacceptable to the end users when the device was powered on. The HomeChoice devices continue to perform as intended, as the issue is only associated with the noise coming from the…Terminated
Dec 8, 2014Z-1039-2015Baxter is issuing a recall for one lot of Integrated Automated Peritoneal Dialysis set with Cassette 3 prong due to complaints received for leakage.Terminated
Oct 21, 2014Z-1031-2015The keypad buttons on HomeChoice devices may be activated without the operator pressing them.Terminated
Oct 6, 2014Z-0939-2015There are additional and updated warning and cautions that are not in the Patient At-Home Guide for HomeChoice Automated PD Systems and HomeChoice PRO Automated PD Systems.Terminated
Apr 3, 2014Z-1245-2015Additional warning in the AMIA APD System labeling for patients vulnerable to unintended Increased Intraperitoneal Volume (IIPV), also known as overfill. The patients include neonates, infants, small children, and any…Terminated
Oct 16, 2013Z-1292-2015The occluder in affected devices 1) may have sharp edges that could cut the pneumatic tubing that connects the occluder blade to the manifold causing an unrecoverable system error, and/or 2) blade may separate from the…Terminated
Oct 7, 2013Z-2609-2016In systems with version 10.4 software Initial Drain logic, the device will attempt to drain the patient to empty at the beginning of therapy (initial drain). The operator can stop, but cannot bypass, the active Initial…Terminated

Other FKX clearances

The newest 510(k)s for the same product code; their median review was 223 days; see every FKX device

Record

Type
Traditional
Received
Oct 4, 2010
Review panel
Gastroenterology, Urology
Regulation
21 CFR 876.5630

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files