uCT 760 with uWS-CT-Dual Energy Analysis, uCT 780 with uWS-CT-Dual Energy Analysis
System, X-Ray, Tomography, Computed. 510(k) K230162, from Shanghai United Imaging Healthcare Co., Ltd. (Shanghai, CN). Product code JAK, class 2.
As of Nov 18, 2026, K230162 is an FDA 510(k) held by Shanghai United Imaging Healthcare Co., Ltd., substantially equivalent on Aug 1, 2023 after 193 days, with 4 recalls tied to it in FDA's recall database.
- Review time: 193 days
- Recalls: 4
- Cited as a predicate by: 3 devices
Indications for use
As the 510(k) summary states them, read from FDA's summary PDFuCT 760/780 is a computed tomography x-ray system intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data taken at different angles and planes and indicated for the whole body (including head, neck, cardiac and vascular).
Predicate devices
The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF- uCT Computed Tomography X-ray SystemK172135 · Shanghai United Imaging Healthcare Co., Ltd. · Mar 2, 2018
- uCT ATLAS with uWS-CT-Dual Energy AnalysisK203448 · Shanghai United Imaging Healthcare Co., Ltd. · Aug 11, 2021
- SOMATOM go.Platform Scanners - SOMATOM go.Now, go.Up, go.All, go.Top, SOMATOM X. Platform Scanners - SOMATOM X.cite and…K211373 · Siemens Medical Solutions USA, Inc. · Aug 27, 2021
Cited as a predicate
3 later 510(k) summaries name this device as a predicate- uCT 780K241079 · Shanghai United Imaging Healthcare Co., Ltd. · Jan 7, 2025
- uOmnispace.CTK233209 · Shanghai United Imaging Healthcare Co., Ltd. · May 17, 2024
- uEXPLORERK231762 · Shanghai United Imaging Healthcare Co., Ltd. · Jan 18, 2024
Recalls
Recalls FDA ties to K230162, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record| Aug 11, 2026 | Z-3279-2026, Z-3280-2026Positron emission tomography and computed tomography x-ray systems have a software problem when Split or Local function is enabled, and multiple RIS exams are merged into a single study, that prevents patient images… | Open |
| Sep 12, 2025 | Z-0220-2026, Z-0219-2026Third-party outer covers can become trapped between patient table and gantry, causing mechanical interference during table movement. | Open |
Other JAK clearances
The newest 510(k)s for the same product code; their median review was 237 days; see every JAK device- AVIEW Aorta MetricsK260391 · Coreline Soft Co., Ltd. · Oct 16, 2026
- CAS-One IRK261225 · Cascination AG · Sep 29, 2026
- BodyTom 64 (NL4100); OmniTom Elite (NL5000)K261125 · Samsung HME America, Inc., dba NeuroLogica Corp. · Sep 22, 2026
- SCENARIA View Phase 6.0K261229 · Fujifilm Corporation · Sep 18, 2026
- FCT iStream Phase 2K253836 · Fujifilm Corporation · Sep 4, 2026
- BodyTom Elite (NL4000)K253450 · NeuroLogica Corp., a Samsung Company · Jul 28, 2026
- CardIQ SuiteK253770 · GE Medical Systems SCS · Jul 21, 2026
- Enhanced Boundary for PCCTK253560 · Ge Medical Systems, LLC · Jun 24, 2026
Record
- Document
- 510(k) summary (PDF) ↗
- Type
- Traditional
- Received
- Jan 20, 2023
- Review panel
- Radiology
- Regulation
- 21 CFR 892.1750
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files