Arca
Skip to content
Tools
Tools

uCT ATLAS with uWS-CT-Dual Energy Analysis

System, X-Ray, Tomography, Computed. 510(k) K203448, from Shanghai United Imaging Healthcare Co., Ltd. (Shanghai, CN). Product code JAK, class 2.

As of Nov 21, 2026, K203448 is an FDA 510(k) held by Shanghai United Imaging Healthcare Co., Ltd., substantially equivalent on Aug 11, 2021 after 261 days, with 3 recalls tied to it in FDA's recall database.

  • Review time: 261 days
  • Recalls: 3
  • Cited as a predicate by: 3 devices

Indications for use

As the 510(k) summary states them, read from FDA's summary PDF

uCT ATLAS is a computed tomography x-ray system, which is intended to produce cross-sectional images of the whole body by computer reconstruction of x-ray transmission data taken at different angles and planes. uCT ATLAS is applicable to head, whole body, cardiac, and vascular x-ray Computed Tomography. uCT ATLAS has the capability to image a whole organ in a single rotation.

Predicate devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a predicate

3 later 510(k) summaries name this device as a predicate

Recalls

Recalls FDA ties to K203448, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
Aug 11, 2026Z-3282-2026, Z-3281-2026Positron emission tomography and computed tomography x-ray systems have a software problem when Split or Local function is enabled, and multiple RIS exams are merged into a single study, that prevents patient images…Open
Sep 12, 2025Z-0221-2026Third-party outer covers can become trapped between patient table and gantry, causing mechanical interference during table movement.Open

Other JAK clearances

The newest 510(k)s for the same product code; their median review was 237 days; see every JAK device

Elsewhere

Record

Type
Traditional
Received
Nov 23, 2020
Review panel
Radiology
Regulation
21 CFR 892.1750

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files