HemoSphere Advanced Monitor, HemoSphere Swan-Ganz Module, HemoSphere Oximetry Cable
Computer, Diagnostic, Programmable. 510(k) K163381, from Edwards Lifesciences, LLC (Irvine, CA). Product code DQK, class 2.
As of Nov 17, 2026, K163381 is an FDA 510(k) held by Edwards Lifesciences, LLC, substantially equivalent on Apr 14, 2017 after 134 days, with 2 recalls tied to it in FDA's recall database.
- Substantially equivalent: Apr 14, 2017
- Review time: 134 days
- Recalls: 2
- Cited as a predicate by: 2 devices
Indications for Use
As the 510(k) summary states them, read from FDA's summary PDFHemoSphere Advanced Monitor with HemoSphere Swan-Ganz Module The HemoSphere Advanced Monitor when used with the HemoSphere Swan-Ganz Module and Edwards Swan-Ganz Catheters is indicated for use in adult and pediatric critical care patients requiring monitoring of cardiac output [continuous (CO) and intermittent (iCO)] and derived hemodynamic parameters in a hospital environment.
Predicate Devices
The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF- Vigilance Ii Continuous Cardiac Output/Oximetry/Volumetric Monitor, Model VIG2K043103 · Edwards Lifesciences, LLC · Dec 9, 2004
- EV1000 Clinical Platform with ClearSight Finger Cuff or ClearSight System, EV1000…K160552 · Edwards Lifesciences, LLC · Jun 1, 2016
Cited as a Predicate
2 later 510(k) summaries name this device as a predicate- HemoSphere Advanced Monitoring Platform, HemoSphere ClearSight ModuleK201446 · Edwards Lifesiences, LLC · Oct 1, 2020
- HemoSphere Advanced Monitor, HemoSphere Pressure CableK180881 · Edwards Lifesciences, LLC · Nov 16, 2018
Recalls
Recalls FDA ties to K163381, newest first; each opens FDA's recall record| Oct 7, 2019 | Z-1928-2020Intermittent communication due to inconsistent crimping of connectors onto cables. | Terminated |
| Oct 2, 2017 | Z-0496-2018Pre-procedural issues related to software defects | Terminated |
Other DQK Clearances
The newest 510(k)s for the same product code; their median review was 147 days; see every DQK device- Vivo (9002)K261886 · Catheter Precision, Inc. · Oct 30, 2026
- PulseAPI SmartWatchK254226 · PulseAI, Ltd. · Oct 27, 2026
- STAR Apollo Mapping SystemK263380 · Rhythm AI, Ltd. · Oct 23, 2026
- Intra-Operative Positioning System (IOPS) and IOPS Visionary SystemK263016 · Centerline Biomedical · Oct 1, 2026
- S-Patch CardioAIK254255 · Wellysis Corp. · Sep 25, 2026
- EnSite X EP SystemK261729 · Abbott Medical · Sep 17, 2026
- AVVIGO+ Multi-Modality Guidance SystemK253692 · Boston Scientific · Aug 21, 2026
- Affera Integrated Mapping SystemK262005 · Medtronic, Inc. · Aug 21, 2026
Record
- Document
- 510(k) summary (PDF) ↗
- Type
- Traditional
- Received
- Dec 1, 2016
- Review panel
- Cardiovascular
- Regulation
- 21 CFR 870.1425
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files