S-Patch CardioAI
Computer, Diagnostic, Programmable. 510(k) K254255, from Wellysis Corp. (Seoul, KR). Product code DQK, class 2.
As of Nov 17, 2026, K254255 is an FDA 510(k) held by Wellysis Corp., substantially equivalent on Sep 25, 2026 after 270 days, with no recalls tied to it in FDA's recall database.
- Substantially equivalent: Sep 25, 2026
- Review time: 270 days
- Recalls: 0
Predicate Devices
The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF- DeepRhythmAIK210822 · Medicalgorithmics S.A. · Jul 27, 2022
Other DQK Clearances
The newest 510(k)s for the same product code; their median review was 147 days; see every DQK device- Vivo (9002)K261886 · Catheter Precision, Inc. · Oct 30, 2026
- PulseAPI SmartWatchK254226 · PulseAI, Ltd. · Oct 27, 2026
- STAR Apollo Mapping SystemK263380 · Rhythm AI, Ltd. · Oct 23, 2026
- Intra-Operative Positioning System (IOPS) and IOPS Visionary SystemK263016 · Centerline Biomedical · Oct 1, 2026
- EnSite X EP SystemK261729 · Abbott Medical · Sep 17, 2026
- AVVIGO+ Multi-Modality Guidance SystemK253692 · Boston Scientific · Aug 21, 2026
- Affera Integrated Mapping SystemK262005 · Medtronic, Inc. · Aug 21, 2026
- CorVista System® with PCWP Add-OnK253576 · Analytics For Life, Inc. · Jul 26, 2026
Record
- Document
- 510(k) summary (PDF) ↗
- Type
- Traditional
- Received
- Dec 29, 2025
- Review panel
- Cardiovascular
- Regulation
- 21 CFR 870.1425
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files