CARDIOSAVE Hybrid Intra-Aortic Balloon Pump

System, Balloon, Intra-Aortic And Control. 510(k) K151254, from Datascope Corp. (Manwah, NJ). Product code DSP, class 2.

As of Nov 27, 2026, K151254 is an FDA 510(k) held by Datascope Corp., substantially equivalent on Jul 2, 2015 after 51 days, with 20 recalls tied to it in FDA's recall database.

  • Review time: 51 days
  • Recalls: 20
  • Cited as a predicate by: 1 device

Indications for use

As the 510(k) summary states them, read from FDA's summary PDF

The CARDIOSAVE Intra-Aortic Balloon Pump is indicated for acute coronary syndrome, cardiac and non-cardiac surgery, or complications of heart failure.

Predicate devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a predicate

1 later 510(k) summary names this device as a predicate

Recalls

Recalls FDA ties to K151254, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
Nov 4, 2025Z-0916-2026The IFU addendum revises the Preventative Maintenance schedule to align with the update introduced in the Service Manual as of June 2023.Open
Oct 23, 2025Z-0932-2026, Z-0931-2026The IFU addendum updates the Vibration and Shock Table to reference the correct standards.Open
Nov 15, 2021Z-0336-2022, Z-0337-2022Firm is initiating a correction due to the possibility of fluid ingress. Fluid entering the device may short various electronic components thus leading to system shutdown, which could result in unsafe hemodynamic…Open
Sep 21, 2021Z-0091-2022A Cardiosave Lithium-Ion Battery Pack used during protocol testing failed to meet the minimum runtime requirement per Getinge internal Product Specification. These nonconforming batteries were inadvertently released to…Open
Sep 8, 2021Z-2615-2021The Cardiosave IABP may unexpectedly shut down when the device is running on AC power, only one battery is installed in the IABP, and the battery is physically removed while the battery is being charged. This occurs…Open
Sep 8, 2021Z-0075-2022The Helium Indicator on the Cardiosave display may over represent the amount of helium in unit in the presence of Electromagnetic Interference.Open
Mar 22, 2021Z-1506-2021, Z-1507-2021There are cybersecurity vulnerabilities in a widely used low-level TCP/IP software library that may result in a loss of communication to the Hospital Information System/Clinical Information System (HIS/CIS).Open
Jun 17, 2019Z-1940-2019, Z-1942-2019If battery maintenance is not performed appropriately, the battery may provide less than the minimum run time of operating power. Revised labeling and training are being provided.Open
Jun 17, 2019Z-1941-2019If battery maintenance is not performed appropriately, the battery may provide less than the minimum run time of operating power. Revised labeling and training are being offered.Open
Jun 17, 2019Z-1943-2019, Z-1944-2019If battery maintenance is not performed per instructions, the battery may provide less than the minimum run time of operating power.Open
Jun 17, 2019Z-1939-2019Battery short-run times and unanticipated stoppage may be due to user not performing battery maintenance and handling as needed. Inadequate labeling is being targeted for this recall, and design updates will be handled…Open
May 3, 2018Z-1928-2018Ingress of fluids into the device can affect various electronic circuit boards which would prevent initiation or continuation of therapy.Open
Jul 17, 2017Z-3087-2017False blood back detection alarm and the ingress of fluids into the IABP affecting various electronic circuit boards could potentially prevent initiation or continuation of therapy.Open
Jun 16, 2017Z-2738-2017The device failed to pump due to an electrical test failure code #58 (power up vent test fail), maintenance code #3, and an autofill failure which has been associated to a patient death due to the failure of the device…Terminated
Dec 23, 2016Z-1386-2017Maquet has received 3 confirmed complaints related to the Li-lon Battery used with the Maquet CARDIOSAVE Hybrid IABP and Maquet CARDIOSAVE Rescue IABP of the battery venting after being accidentally dropped.Terminated

Other DSP clearances

The newest 510(k)s for the same product code; see every DSP device

Record

Type
Special
Received
May 12, 2015
Review panel
Cardiovascular
Regulation
21 CFR 870.3535

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files