Cardiosave Hybrid Intra-Aortic Balloon Pump

System, Balloon, Intra-Aortic And Control. 510(k) K112372, from Cardiac Assist, Maquet Cardiovascular, LLC (Manwah, NJ). Product code DSP, class 2.

As of Nov 27, 2026, K112372 is an FDA 510(k) held by Cardiac Assist, Maquet Cardiovascular, LLC, substantially equivalent on Sep 15, 2011 after 29 days, with 25 recalls tied to it in FDA's recall database.

  • Review time: 29 days
  • Recalls: 25+
  • Cited as a predicate by: 2 devices

Indications for use

As the 510(k) summary states them, read from FDA's summary PDF

The CARDIOSAVE™ Intra Aortic Balloon Pump is an electromechanical system used to inflate and deflate intra-aortic balloons. It provides temporary support to the left ventricle via the principle of counterpulsation. The intra-aortic balloon is placed in the descending aorta, just distal to the left subclavian artery.

Predicate devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a predicate

2 later 510(k) summaries name this device as a predicate

Recalls

Recalls FDA ties to K112372, newest first; the 25 newest; recall numbers filed together for one problem share a row, and each opens FDA's recall record
Feb 6, 2026Z-1587-2026The Battery Charging Station is an optional dual-bay charger used to charge the Cardiosave IABP Lithium- Ion batteries when they are not being used to operate the IABP. Getinge has identified that a protruding screw in…Open
Nov 4, 2025Z-0916-2026The IFU addendum revises the Preventative Maintenance schedule to align with the update introduced in the Service Manual as of June 2023.Open
Oct 23, 2025Z-0932-2026, Z-0931-2026The IFU addendum updates the Vibration and Shock Table to reference the correct standards.Open
May 4, 2023Z-1907-2023The digital IFUs provided with the Cardiosave IABP during the software update to version D.00 incorrectly annotated compliance to standard 60601-1-12:2014 within the updated IFU.Open
Apr 4, 2023Z-1462-2023The power cord supplied (Type J plug end) for the Cardiosave Hybrid IABP to Brazil is incompatible with the Brazilian Type N power receptacle.Open
Feb 28, 2023Z-1276-2023, Z-1277-2023, Z-1278-2023, Z-1279-2023There is an issue related to the Safety Disk not meeting a performance requirement that may impact Intra-Aortic Balloon catheter (IAB) displacement volume. This potential failure of an impacted Safety Disk may lead to…Open
Feb 7, 2023Z-1229-2023, Z-1230-2023There have been reported failures of the high pressure helium regulator, which may cause a helium leak in the Cardiosave Hospital Cart. In instance of helium regulator failure, a Pump Console s internal reservoir of…Open
Feb 7, 2023Z-1146-2023, Z-1145-2023An unexpected shutdown of the IABP may occur due to a failure of the connection between the Coiled Cord cable and the Cable Assembly backplane to the Coiled Cord cable which provides the communication between the…Open
Nov 15, 2021Z-0336-2022, Z-0337-2022Firm is initiating a correction due to the possibility of fluid ingress. Fluid entering the device may short various electronic components thus leading to system shutdown, which could result in unsafe hemodynamic…Open
Sep 21, 2021Z-0091-2022A Cardiosave Lithium-Ion Battery Pack used during protocol testing failed to meet the minimum runtime requirement per Getinge internal Product Specification. These nonconforming batteries were inadvertently released to…Open
Sep 8, 2021Z-2615-2021The Cardiosave IABP may unexpectedly shut down when the device is running on AC power, only one battery is installed in the IABP, and the battery is physically removed while the battery is being charged. This occurs…Open
Sep 8, 2021Z-0075-2022The Helium Indicator on the Cardiosave display may over represent the amount of helium in unit in the presence of Electromagnetic Interference.Open
Mar 22, 2021Z-1506-2021, Z-1507-2021There are cybersecurity vulnerabilities in a widely used low-level TCP/IP software library that may result in a loss of communication to the Hospital Information System/Clinical Information System (HIS/CIS).Open
Jun 17, 2019Z-1940-2019, Z-1942-2019If battery maintenance is not performed appropriately, the battery may provide less than the minimum run time of operating power. Revised labeling and training are being provided.Open
Jun 17, 2019Z-1941-2019If battery maintenance is not performed appropriately, the battery may provide less than the minimum run time of operating power. Revised labeling and training are being offered.Open
Jun 17, 2019Z-1939-2019Battery short-run times and unanticipated stoppage may be due to user not performing battery maintenance and handling as needed. Inadequate labeling is being targeted for this recall, and design updates will be handled…Open

Other DSP clearances

The newest 510(k)s for the same product code; see every DSP device

Record

Type
Special
Received
Aug 17, 2011
Review panel
Cardiovascular
Regulation
21 CFR 870.3535

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files