Cardiosave Hybrid Intra-Aortic Balloon Pump
System, Balloon, Intra-Aortic And Control. 510(k) K112372, from Cardiac Assist, Maquet Cardiovascular, LLC (Manwah, NJ). Product code DSP, class 2.
As of Nov 27, 2026, K112372 is an FDA 510(k) held by Cardiac Assist, Maquet Cardiovascular, LLC, substantially equivalent on Sep 15, 2011 after 29 days, with 25 recalls tied to it in FDA's recall database.
- Review time: 29 days
- Recalls: 25+
- Cited as a predicate by: 2 devices
The CARDIOSAVE™ Intra Aortic Balloon Pump is an electromechanical system used to inflate and deflate intra-aortic balloons. It provides temporary support to the left ventricle via the principle of counterpulsation. The intra-aortic balloon is placed in the descending aorta, just distal to the left subclavian artery.
- Datascope Fiber Optic Pressure Sensor Intra-Aortic Balloon Catheter and CS300 Intra-Aortic Balloon Pump Control SystemK063525 · Datascope Corp. · Jan 5, 2007
- CARDIOSAVE Intra-Aortic Balloon PumpK181122 · Datascope Corp. · May 31, 2018
- CARDIOSAVE Hybrid Intra-Aortic Balloon PumpK151254 · Datascope Corp. · Jul 2, 2015
| Feb 6, 2026 | Z-1587-2026The Battery Charging Station is an optional dual-bay charger used to charge the Cardiosave IABP Lithium- Ion batteries when they are not being used to operate the IABP. Getinge has identified that a protruding screw in… | Open |
| Nov 4, 2025 | Z-0916-2026The IFU addendum revises the Preventative Maintenance schedule to align with the update introduced in the Service Manual as of June 2023. | Open |
| Oct 23, 2025 | Z-0932-2026, Z-0931-2026The IFU addendum updates the Vibration and Shock Table to reference the correct standards. | Open |
| May 4, 2023 | Z-1907-2023The digital IFUs provided with the Cardiosave IABP during the software update to version D.00 incorrectly annotated compliance to standard 60601-1-12:2014 within the updated IFU. | Open |
| Apr 4, 2023 | Z-1462-2023The power cord supplied (Type J plug end) for the Cardiosave Hybrid IABP to Brazil is incompatible with the Brazilian Type N power receptacle. | Open |
| Feb 28, 2023 | Z-1276-2023, Z-1277-2023, Z-1278-2023, Z-1279-2023There is an issue related to the Safety Disk not meeting a performance requirement that may impact Intra-Aortic Balloon catheter (IAB) displacement volume. This potential failure of an impacted Safety Disk may lead to… | Open |
| Feb 7, 2023 | Z-1229-2023, Z-1230-2023There have been reported failures of the high pressure helium regulator, which may cause a helium leak in the Cardiosave Hospital Cart. In instance of helium regulator failure, a Pump Console s internal reservoir of… | Open |
| Feb 7, 2023 | Z-1146-2023, Z-1145-2023An unexpected shutdown of the IABP may occur due to a failure of the connection between the Coiled Cord cable and the Cable Assembly backplane to the Coiled Cord cable which provides the communication between the… | Open |
| Nov 15, 2021 | Z-0336-2022, Z-0337-2022Firm is initiating a correction due to the possibility of fluid ingress. Fluid entering the device may short various electronic components thus leading to system shutdown, which could result in unsafe hemodynamic… | Open |
| Sep 21, 2021 | Z-0091-2022A Cardiosave Lithium-Ion Battery Pack used during protocol testing failed to meet the minimum runtime requirement per Getinge internal Product Specification. These nonconforming batteries were inadvertently released to… | Open |
| Sep 8, 2021 | Z-2615-2021The Cardiosave IABP may unexpectedly shut down when the device is running on AC power, only one battery is installed in the IABP, and the battery is physically removed while the battery is being charged. This occurs… | Open |
| Sep 8, 2021 | Z-0075-2022The Helium Indicator on the Cardiosave display may over represent the amount of helium in unit in the presence of Electromagnetic Interference. | Open |
| Mar 22, 2021 | Z-1506-2021, Z-1507-2021There are cybersecurity vulnerabilities in a widely used low-level TCP/IP software library that may result in a loss of communication to the Hospital Information System/Clinical Information System (HIS/CIS). | Open |
| Jun 17, 2019 | Z-1940-2019, Z-1942-2019If battery maintenance is not performed appropriately, the battery may provide less than the minimum run time of operating power. Revised labeling and training are being provided. | Open |
| Jun 17, 2019 | Z-1941-2019If battery maintenance is not performed appropriately, the battery may provide less than the minimum run time of operating power. Revised labeling and training are being offered. | Open |
| Jun 17, 2019 | Z-1939-2019Battery short-run times and unanticipated stoppage may be due to user not performing battery maintenance and handling as needed. Inadequate labeling is being targeted for this recall, and design updates will be handled… | Open |
- AC3™ Range™ Intra-Aortic Balloon PumpK250542 · Arrow International, LLC · Mar 26, 2025
- AC3™ Series IABPK232343 · Arrow International, LLC · Aug 30, 2023
- Document
- 510(k) summary (PDF) ↗
- Type
- Special
- Received
- Aug 17, 2011
- Review panel
- Cardiovascular
- Regulation
- 21 CFR 870.3535
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files