CARDIOSAVE Intra-Aortic Balloon Pump
System, Balloon, Intra-Aortic And Control. 510(k) K181122, from Datascope Corp. (Manwah, NJ). Product code DSP, class 2.
As of Nov 27, 2026, K181122 is an FDA 510(k) held by Datascope Corp., substantially equivalent on May 31, 2018 after 31 days, with 25 recalls tied to it in FDA's recall database.
- Review time: 31 days
- Recalls: 25+
The CARDIOSAVE IABP (Hybrid Model) is indicated for acute coronary syndrome, cardiac and non-cardiac surgery, or complications of heart failure.
- Cardiosave Hybrid Intra-Aortic Balloon PumpK112372 · Cardiac Assist, Maquet Cardiovascular, LLC · Sep 15, 2011
| Nov 4, 2025 | Z-0916-2026The IFU addendum revises the Preventative Maintenance schedule to align with the update introduced in the Service Manual as of June 2023. | Open |
| Oct 23, 2025 | Z-0932-2026, Z-0931-2026The IFU addendum updates the Vibration and Shock Table to reference the correct standards. | Open |
| Aug 8, 2024 | Z-3134-2024, Z-3135-2024Firm has developed a software correction (Revision D.00) outside the US to mitigate product problems that pose a risk of hemodynamic instability and failure of transmission of clinical data in Cardiosave Hybrid and… | Open |
| Aug 8, 2024 | Z-3137-2024, Z-3136-2024Firm has developed a software correction (Revision D.01) for outside the US to mitigate product problems that pose a risk of hemodynamic instability and failure of transmission of clinical data in Cardiosave Hybrid and… | Open |
| Jul 31, 2023 | Z-2486-2023, Z-2487-2023Users were experiencing a failure in the IAB Fiber Optic Sensor input on the IABP when inserting the Intra-Aortic Balloon Fiber Optic connector. | Open |
| Jul 31, 2023 | Z-2433-2023, Z-2434-2023Users reported "System Over Temperature" alarms associated with a loss of pumping and/or the Cardiosave system entering Standby mode. | Open |
| Jul 31, 2023 | Z-2432-2023, Z-2431-2023Users were reporting that the device was not charging as expected. It was discovered that users were unaware that the Cardiosave console was not completely inserted into the hospital cart. If the console is not fully… | Open |
| Jul 31, 2023 | Z-2438-2023, Z-2437-2023Users were identifying autofill failure conditions on the devices causing pump stops. | Open |
| Jul 31, 2023 | Z-2436-2023, Z-2435-2023Users were reporting instances of "Gas Loss in IAB Circuit" and "Gas Gain in IAB Circuit" alarms while providing therapy. An internal investigation of the complaints determined that there is a potential trigger for… | Open |
| Jun 5, 2023 | Z-2149-2023, Z-2150-2023IABP may lose the ability to charge batteries in one or both bay slots. Therapy may be interrupted if batteries fail to charge and the device is disconnected from AC power. Low battery alarms may alert the User to the… | Open |
| Jun 5, 2023 | Z-2151-2023, Z-2152-2023An unexpected shutdown of the IABP may occur due to failures of tantalum capacitors in the Power Management Board and/or Solenoid Board. This issue may lead to an unexpected interruption of therapy. | Open |
| Feb 7, 2023 | Z-1232-2023, Z-1231-2023Firm has received reports of damaged, worn, or torn O-rings on the Cardiosave Pump Console quick disconnect fitting, which results in helium tank leaks. If the helium supply is depleted, therapy will be interrupted. | Open |
| Feb 7, 2023 | Z-1147-2023, Z-1148-2023An unexpected shutdown of the IABP may occur due to loss of communication between the Executive Processor PCBA and the Video Generator PCBA. This may result in an interruption to therapy which may threaten the… | Open |
- AC3™ Range™ Intra-Aortic Balloon PumpK250542 · Arrow International, LLC · Mar 26, 2025
- AC3™ Series IABPK232343 · Arrow International, LLC · Aug 30, 2023
- Document
- 510(k) summary (PDF) ↗
- Type
- Special
- Received
- Apr 30, 2018
- Review panel
- Cardiovascular
- Regulation
- 21 CFR 870.3535
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files