CARDIOSAVE Hybrid and Rescue Intra-Aortic Balloon Pump, CARDIOSAVE Li-Ion Battery Transport and Storage Case
System, Balloon, Intra-Aortic And Control. 510(k) K163542, from Datascope Corp. (Manwah, NJ). Product code DSP, class 2.
As of Nov 27, 2026, K163542 is an FDA 510(k) held by Datascope Corp., substantially equivalent on Jan 31, 2017 after 46 days, with 14 recalls tied to it in FDA's recall database.
- Review time: 46 days
- Recalls: 14
The CARDIOSAVE Intra-Aortic Balloon Pump is indicated for acute coronary syndrome, cardiac and non-cardiac surgery, or complications of heart failure. COMPARISON OF
- CARDIOSAVE Hybrid Intra-Aortic Balloon PumpK151254 · Datascope Corp. · Jul 2, 2015
| Nov 4, 2025 | Z-0917-2026The IFU addendum revises the Preventative Maintenance schedule to align with the update introduced in the Service Manual as of June 2023. | Open |
| Oct 23, 2025 | Z-0932-2026, Z-0931-2026The IFU addendum updates the Vibration and Shock Table to reference the correct standards. | Open |
| Nov 15, 2021 | Z-0336-2022, Z-0337-2022Firm is initiating a correction due to the possibility of fluid ingress. Fluid entering the device may short various electronic components thus leading to system shutdown, which could result in unsafe hemodynamic… | Open |
| Sep 21, 2021 | Z-0091-2022A Cardiosave Lithium-Ion Battery Pack used during protocol testing failed to meet the minimum runtime requirement per Getinge internal Product Specification. These nonconforming batteries were inadvertently released to… | Open |
| Sep 8, 2021 | Z-2615-2021The Cardiosave IABP may unexpectedly shut down when the device is running on AC power, only one battery is installed in the IABP, and the battery is physically removed while the battery is being charged. This occurs… | Open |
| Sep 8, 2021 | Z-0075-2022The Helium Indicator on the Cardiosave display may over represent the amount of helium in unit in the presence of Electromagnetic Interference. | Open |
| Mar 22, 2021 | Z-1506-2021, Z-1507-2021There are cybersecurity vulnerabilities in a widely used low-level TCP/IP software library that may result in a loss of communication to the Hospital Information System/Clinical Information System (HIS/CIS). | Open |
| Jun 17, 2019 | Z-1940-2019, Z-1942-2019If battery maintenance is not performed appropriately, the battery may provide less than the minimum run time of operating power. Revised labeling and training are being provided. | Open |
| Jun 17, 2019 | Z-1941-2019If battery maintenance is not performed appropriately, the battery may provide less than the minimum run time of operating power. Revised labeling and training are being offered. | Open |
| Jun 17, 2019 | Z-1939-2019Battery short-run times and unanticipated stoppage may be due to user not performing battery maintenance and handling as needed. Inadequate labeling is being targeted for this recall, and design updates will be handled… | Open |
- AC3™ Range™ Intra-Aortic Balloon PumpK250542 · Arrow International, LLC · Mar 26, 2025
- AC3™ Series IABPK232343 · Arrow International, LLC · Aug 30, 2023
- Document
- 510(k) summary (PDF) ↗
- Type
- Special
- Received
- Dec 16, 2016
- Review panel
- Cardiovascular
- Regulation
- 21 CFR 870.3535
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files