Evis Exera Ii Duodenovideoscope Olympus Tjf Type Q180V
Duodenoscope And Accessories, Flexible/Rigid. 510(k) K143153, from Olympus Medical Systems Corp. (Hachioji-Shi, JP). Product code FDT, class 2.
As of Nov 18, 2026, K143153 is an FDA 510(k) held by Olympus Medical Systems Corp., substantially equivalent on Jan 15, 2016 after 438 days, with 3 recalls tied to it in FDA's recall database.
- Substantially equivalent: Jan 15, 2016
- Review time: 438 days
- Recalls: 3
- Cited as a predicate by: 5 devices
Indications for Use
As the 510(k) summary states them, read from FDA's summary PDFThis instrument has been designed to be used with an Olympus video system center, light source, documentation equipment, monitor, EndoTherapy accessories (such as a biopsy forceps), and other ancillary equipment for endoscopy and endoscopic surgery within the duodenum.
Predicate Devices
The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF- Xtjf Type Q160VF1K080403 · Olympus Medical Systems Corporation · May 20, 2008
Cited as a Predicate
5 later 510(k) summaries name this device as a predicate- Ambu® aScope™ Duodeno 2, Ambu® aBox™ 2K233886 · Ambu A/S · Apr 16, 2024
- Ambu Duodeno SystemK201098 · Ambu Innovation GmbH · Jul 17, 2020
- EXALT Model D Single-Use Duodenoscope, EXALT ControllerK193202 · Boston Scientific Corporation · Dec 13, 2019
- Evis Exera III Duodenovideoscope Olympus TJF-Q190VK193182 · Olympus Medical Systems Corp. · Jan 17, 2020
- FUJIFILM Duodenoscope ModelK181745 · Fujifilm Corporation · Mar 28, 2019
Recalls
Recalls FDA ties to K143153, newest first; each opens FDA's recall record| Sep 25, 2023 | Z-0346-2024Channel air drying process was not validated, and that a small percentage of scopes returned to customers after repair had a wet channel. Water remaining in the channel of the endoscopes has the potential for… | Open, Classified |
| Nov 2, 2021 | Z-0379-2022Use of a TJF-Q180V with adhesive deterioration or other damages may pose a risk of endoscope contamination due to ineffective reprocessing or fluid invasion; contaminated endoscope can present an infection risk to… | Open, Classified |
| Dec 17, 2018 | Z-1010-2019The adhesive used in the repair of the endoscope was incorrectly prepared. It is unknown what affect the nonconforming adhesive mixture would have on the durability or effectiveness of the gluing operation for the… | Terminated |
Other FDT Clearances
The newest 510(k)s for the same product code; their median review was 94 days; see every FDT device- FUJIFILM Endoscope Model ED-S100TP; FUJIFILM Endoscope Model ED-S100XP; FUJIFILM Processor Model VS-1000K251861 · Fujifilm Corporation · Mar 9, 2026
- Single Use Distal Cover MAJ-2315 (MAJ-2315)K253646 · Olympus Medical Systems Corp. · Feb 22, 2026
- Evis Exera Iii Duodenovideoscope Olympus TJF-Q190V (TJF-Q190V)K251867 · Olympus Medical Systems Corp. · Sep 19, 2025
- Evis Exera III Duodenovideoscope Olympus TJF-Q190V (TJF-Q190V)K250701 · Olympus Medical Systems Corp. · Jun 5, 2025
- PENTAX Medical Video Duodenoscope (ED34-i10T2s); PENTAX Medical Video Processor (EPK-i8020c); PENTAX Medical Single…K233942 · Pentax of America, Inc. · Aug 1, 2024
- Ambu® aScope Duodeno 2, Ambu® aBox 2K233886 · Ambu A/S · Apr 16, 2024
- Evis Exera III Duodenovideoscope Olympus TJF-Q190VK220587 · Olympus Medical Systems Corp. · May 11, 2022
- PENTAX Medical Video Duedenoscope ED34-i10T2K210710 · Pentax of America, Inc. · Apr 9, 2021
Record
- Document
- 510(k) summary (PDF) ↗
- Type
- Traditional
- Received
- Nov 3, 2014
- Review panel
- Gastroenterology, Urology
- Regulation
- 21 CFR 876.1500
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files