Arca
Skip to content
Tools
Tools

EXALT Model D Single-Use Duodenoscope, EXALT Controller

Duodenoscope And Accessories, Flexible/Rigid. 510(k) K193202, from Boston Scientific Corporation (Marlboro, MA). Product code FDT, class 2.

As of Nov 18, 2026, K193202 is an FDA 510(k) held by Boston Scientific Corporation, substantially equivalent on Dec 13, 2019 after 23 days, with 6 recalls tied to it in FDA's recall database.

  • Substantially equivalent: Dec 13, 2019
  • Review time: 23 days
  • Recalls: 6
  • Cited as a predicate by: 1 device

Indications for Use

As the 510(k) summary states them, read from FDA's summary PDF

EXALT Model D Single Use Duodenoscope: The EXALT Model D Single-Use Duodenoscope is intended for use with a Boston Scientific endoscopic video imaging system, for endoscopy and endoscopic surgery within the duodenum. EXALT Controller: The EXALT Controller is intended for use with a Boston Scientific endoscope for endoscopic diagnosis, treatment, and video observation.

Predicate Devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a Predicate

1 later 510(k) summary names this device as a predicate

Recalls

Recalls FDA ties to K193202, newest first; each opens FDA's recall record
Sep 27, 2023Z-0396-2024Poor image quality due to fluid ingress in the lens.Open, Classified
Sep 27, 2023Z-0397-2024Poor image quality due to fluid ingress in the lens.Open, Classified
Mar 3, 2022Z-1081-2022Updated Instructions for Use (IFU) for EXALT Model D Single-Use Duodenoscope highlight the known risk of perforation and provide best clinical practices related to insertion, advancement, and removal of this device.Open, Classified
Mar 3, 2022Z-1079-2022Updated Instructions for Use (IFU) for EXALT Model D Single-Use Duodenoscope highlight the known risk of perforation and provide best clinical practices related to insertion, advancement, and removal of this device.Open, Classified
Mar 3, 2022Z-1082-2022Updated Instructions for Use (IFU) for EXALT Model D Single-Use Duodenoscope highlight the known risk of perforation and provide best clinical practices related to insertion, advancement, and removal of this device.Open, Classified
Mar 3, 2022Z-1080-2022Updated Instructions for Use (IFU) for EXALT Model D Single-Use Duodenoscope highlight the known risk of perforation and provide best clinical practices related to insertion, advancement, and removal of this device.Open, Classified

Other FDT Clearances

The newest 510(k)s for the same product code; their median review was 94 days; see every FDT device

Record

Type
Traditional
Received
Nov 20, 2019
Review panel
Gastroenterology, Urology
Regulation
21 CFR 876.1500

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files