Evis Exera III Duodenovideoscope Olympus TJF-Q190V
Duodenoscope And Accessories, Flexible/Rigid. 510(k) K193182, from Olympus Medical Systems Corp. (Hachioji-Shi, JP). Product code FDT, class 2.
As of Nov 18, 2026, K193182 is an FDA 510(k) held by Olympus Medical Systems Corp., substantially equivalent on Jan 17, 2020 after 60 days, with 4 recalls tied to it in FDA's recall database.
- Substantially equivalent: Jan 17, 2020
- Review time: 60 days
- Recalls: 4
- Cited as a predicate by: 1 device
Indications for Use
As the 510(k) summary states them, read from FDA's summary PDFThis instrument has been designed to be used with an Olympus video system center, light source, documentation equipment, monitor, EndoTherapy accessories (such as a biopsy forceps), and other ancillary equipment for endoscopy and endoscopic surgery within the duodenum.
Predicate Devices
The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF- Evis Exera Ii Duodenovideoscope Olympus Tjf Type Q180VK143153 · Olympus Medical Systems Corp. · Jan 15, 2016
Cited as a Predicate
1 later 510(k) summary names this device as a predicate- Evis Exera III Duodenovideoscope Olympus TJF-Q190VK202661 · Olympus Medical Systems Corp. · Dec 8, 2020
Recalls
Recalls FDA ties to K193182, newest first; each opens FDA's recall record| Sep 25, 2023 | Z-0346-2024Channel air drying process was not validated, and that a small percentage of scopes returned to customers after repair had a wet channel. Water remaining in the channel of the endoscopes has the potential for… | Open, Classified |
| Jul 27, 2023 | Z-2467-2023Recent reports of patient infection. | Open, Classified |
| Apr 18, 2023 | Z-1613-2023The Single Use Distal Cover MAJ-2315 may cause mucosal injury and/or become detached from the endoscope during use. Consequences of a detached cover could include the risk of aspiration, inhalation, or obstruction that… | Open, Classified |
| Jan 28, 2021 | Z-1292-2021Olympus Medical Systems Corporation (OMSC) has received complaints of mucosal injury associated with the forceps elevator region of the TJF-Q190V duodenoscope, including the distal cover (model MAJ- 2315). | Terminated |
Other FDT Clearances
The newest 510(k)s for the same product code; their median review was 94 days; see every FDT device- FUJIFILM Endoscope Model ED-S100TP; FUJIFILM Endoscope Model ED-S100XP; FUJIFILM Processor Model VS-1000K251861 · Fujifilm Corporation · Mar 9, 2026
- Single Use Distal Cover MAJ-2315 (MAJ-2315)K253646 · Olympus Medical Systems Corp. · Feb 22, 2026
- Evis Exera Iii Duodenovideoscope Olympus TJF-Q190V (TJF-Q190V)K251867 · Olympus Medical Systems Corp. · Sep 19, 2025
- Evis Exera III Duodenovideoscope Olympus TJF-Q190V (TJF-Q190V)K250701 · Olympus Medical Systems Corp. · Jun 5, 2025
- PENTAX Medical Video Duodenoscope (ED34-i10T2s); PENTAX Medical Video Processor (EPK-i8020c); PENTAX Medical Single…K233942 · Pentax of America, Inc. · Aug 1, 2024
- Ambu® aScope Duodeno 2, Ambu® aBox 2K233886 · Ambu A/S · Apr 16, 2024
- Evis Exera III Duodenovideoscope Olympus TJF-Q190VK220587 · Olympus Medical Systems Corp. · May 11, 2022
- PENTAX Medical Video Duedenoscope ED34-i10T2K210710 · Pentax of America, Inc. · Apr 9, 2021
Record
- Document
- 510(k) summary (PDF) ↗
- Type
- Traditional
- Received
- Nov 18, 2019
- Review panel
- Gastroenterology, Urology
- Regulation
- 21 CFR 876.1500
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files