STERRAD NX Sterilizer, STERRAD 100NX Sterilizer

Sterilizer, Chemical. 510(k) K142454, from Advanced Sterilization Products (Irvin, CA). Product code MLR, class 2.

As of Nov 24, 2026, K142454 is an FDA 510(k) held by Advanced Sterilization Products, substantially equivalent on Apr 3, 2015 after 213 days, with 2 recalls tied to it in FDA's recall database.

  • Review time: 213 days
  • Recalls: 2
  • Cited as a predicate by: 4 devices

Indications for use

As the 510(k) summary states them, read from FDA's summary PDF

FOR FDA USE ONLY Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)

Predicate devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a predicate

4 later 510(k) summaries name this device as a predicate

Recalls

Recalls FDA ties to K142454, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
Jan 9, 2017Z-1197-2017Advanced Sterilization Products (ASP) has identified that in a rare sequence of events following an external power outage, the system door interlock can be become disengaged during system reboot for a 20-second duration.Terminated
Jul 17, 2015Z-2375-2015Advanced Sterilization PRoducts (ASP) is recalling the Wall Chart for STERRAD 100NX System because it incorrectly states "APTIMAX Instrument Trays can be enclosed in either polypropylene wrap or in Tyvek Pouches with…Terminated

Other MLR clearances

The newest 510(k)s for the same product code; their median review was 216 days; see every MLR device

Record

Type
Traditional
Received
Sep 2, 2014
Review panel
General Hospital
Regulation
21 CFR 880.6860

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files