Ge Datex-Ohmeda Aisys CS2 (Aisys CS^2)
Gas-Machine, Anesthesia. 510(k) K132530, from Datex-Ohmeda, Inc. (Madison, WI). Product code BSZ, class 2.
As of Nov 18, 2026, K132530 is an FDA 510(k) held by Datex-Ohmeda, Inc., substantially equivalent on Dec 20, 2013 after 130 days, with 3 recalls tied to it in FDA's recall database.
- Review time: 130 days
- Recalls: 3
- Cited as a predicate by: 2 devices
Indications for use
As the 510(k) summary states them, read from FDA's summary PDFThe GE Datex-Ohmeda Aisys CS² Anesthesia System is intended to provide general inhalation anesthesia and ventilatory support to a wide range of patients (neonatal, pediatric, adult). The device is intended for volume or pressure control ventilation.
Predicate devices
The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF- Ge Datex-Ohmeda AisysK110213 · Datex-Ohmeda, Inc. · Jul 20, 2011
Cited as a predicate
2 later 510(k) summaries name this device as a predicate- Aestiva 7900, Aestiva MRI, Aestiva 7100, Aestiva 7100 Compact, Aespire 100, Aespire 7100, Aespire 7900, Aespire View,…K172045 · Datex-Ohmeda, Inc. · Nov 3, 2017
- Aisys CS2K170872 · Datex-Ohmeda, Inc. · Aug 15, 2017
Recalls
Recalls FDA ties to K132530, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record| Mar 15, 2016 | Z-1562-2016The vacuum suction tubing could be kinked within the machine resulting in suction less than the 20 lpm required by ISO 10079-3. This could result in inadequate suction performance to clear the airway and inadequate… | Terminated |
| Feb 17, 2016 | Z-1381-2016GE Healthcare has recently become aware of a potential safety issue involving a momentary, self-correcting anesthetic agent bolus when using 21% oxygen on all Aisys CS2 and upgraded Aisys anesthesia devices. | Terminated |
| Dec 15, 2015 | Z-0681-2016The Drive Gas Check Valve of certain Anesthesia products could become stuck in a fixed open position, which could cause pressure to build in the mechanical ventilation cycle. If this issue is left unresolved, it could… | Terminated |
Other BSZ clearances
The newest 510(k)s for the same product code; their median review was 265 days; see every BSZ device- A5 Anesthesia System (A5); A7 Anesthesia System (A7)K253330 · Shenzhen Mindray Bio-Medical Electronics Co., Ltd. · Jun 25, 2026
- Atlan (A100); Atlan (A100 XL); Atlan(A350); Atlan (A350 XL)K252746 · Drägerwerk AG Co. Kgaa · May 21, 2026
- Codonics Safe Labeling System-Point of Care Station (PCS) (SLS-700i and related 610i, 620i, 630i series enabled to…K251352 · Codonics.Incorporated · Jul 11, 2025
- AtlanK230931 · Drägerwerk AG Co. Kgaa · Jul 23, 2023
- Carestation 750/750cK213867 · Datex-Ohmeda, Inc. · Mar 28, 2023
- A8, A9 Anesthesia SystemK201957 · Shenzhen Mindray Bio-Medical Electronics Co., Ltd. · Mar 26, 2021
Record
- Document
- 510(k) summary (PDF) ↗
- Type
- Traditional
- Received
- Aug 12, 2013
- Review panel
- Anesthesiology
- Regulation
- 21 CFR 868.5160
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files