Ge Datex-Ohmeda Aisys

Gas-Machine, Anesthesia. 510(k) K110213, from Datex-Ohmeda, Inc. (Madison, WI). Product code BSZ, class 2.

As of Nov 18, 2026, K110213 is an FDA 510(k) held by Datex-Ohmeda, Inc., substantially equivalent on Jul 20, 2011 after 176 days, with 9 recalls tied to it in FDA's recall database.

  • Review time: 176 days
  • Recalls: 9
  • Cited as a predicate by: 3 devices

Indications for use

As the 510(k) summary states them, read from FDA's summary PDF

The GE Datex-Ohmeda Aisys Anesthesia System is intended to provide general inhalation anesthesia and ventilatory support to a wide range of patients (neonatal, pediatric, adult). The device is intended for volume or pressure control ventilation. The Aisys is not suitable for use in a MRI environment.

Predicate devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a predicate

3 later 510(k) summaries name this device as a predicate

Recalls

Recalls FDA ties to K110213, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
Sep 27, 2018Z-0411-2019, Z-0413-2019, Z-0412-2019Upgraded Aisys anesthesia devices and Aisys CS2 anesthesia devices with PSVPro Spontaneous Breathing Modes do not provide the PSVPro ventilation mode feature of transitioning to SIMV PCV backup ventilation mode. PSVPro…Terminated
Mar 15, 2016Z-1562-2016The vacuum suction tubing could be kinked within the machine resulting in suction less than the 20 lpm required by ISO 10079-3. This could result in inadequate suction performance to clear the airway and inadequate…Terminated
Feb 17, 2016Z-1382-2016GE Healthcare has recently become aware of a potential safety issue involving a momentary, self-correcting anesthetic agent bolus when using 21% oxygen on all Aisys CS2 and upgraded Aisys anesthesia devices.Terminated
Dec 15, 2015Z-0680-2016The Drive Gas Check Valve of certain Anesthesia products could become stuck in a fixed open position, which could cause pressure to build in the mechanical ventilation cycle. If this issue is left unresolved, it could…Terminated
Sep 16, 2013Z-0009-2014GE Healthcare has recently become aware of a potential safety issue involving the Avance, Avance CS2, and Aisys anesthesia devices. While operating in Pressure Control Ventilation Volume Guarantee (PCV-VG) mode, the…Terminated
Aug 28, 2013Z-0010-2014GE Healthcare has recently become aware of a potential safety issue involving unresponsive buttons on the display of the Engstrm ventilator and the Aespire View, Aisys, and Avance anesthesia machines. As a result of a…Terminated
May 14, 2013Z-1443-2013GE Healthcare has recently become aware of a potential safety issue involving the reusable CO2 absorbent canister accessory used with Aespire, Aespire View, Avance, Avance CS2, Aisys, ADU, 9100, M900, M904, 9300…Terminated

Other BSZ clearances

The newest 510(k)s for the same product code; their median review was 265 days; see every BSZ device

Record

Type
Traditional
Received
Jan 25, 2011
Review panel
Anesthesiology
Regulation
21 CFR 868.5160

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files