Arca
Skip to content

Servo-I Ventilator System

Ventilator, Continuous, Facility Use. 510(k) K123149, from Maquet Critical Care AB (Wayne, NJ). Product code CBK, class 2.

As of Nov 17, 2026, K123149 is an FDA 510(k) held by Maquet Critical Care AB, substantially equivalent on Jun 20, 2014 after 623 days, with 2 recalls tied to it in FDA's recall database.

  • Substantially equivalent: Jun 20, 2014
  • Review time: 623 days
  • Recalls: 2
  • Cited as a predicate by: 3 devices

Indications for Use

As the 510(k) summary states them, read from FDA's summary PDF

The SERVO-i Ventilator System is intended for treatment and monitoring of patients in the range of neonates, infants, and adults with respiratory failure or respiratory insufficiency. SERVO-i is a ventilator system to be used only by healthcare providers in hospitals or healthcare facilities and for in-hospital transport.

Predicate Devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a Predicate

3 later 510(k) summaries name this device as a predicate

Recalls

Recalls FDA ties to K123149, newest first; each opens FDA's recall record
Sep 17, 2020Z-0092-2021A potentially shorter than specified nebulizer connector may result in difficulty in installing the Aeroneb Nebulizer module to the Servo-i device successfully.Open, Classified
Jan 12, 2015Z-1224-2015A calibration problem was detected in a specific number of the Servo-I ventilator system CO2 analyzer models that may have been caused by the module's internal circuit board function.Terminated

Other CBK Clearances

The newest 510(k)s for the same product code; their median review was 235 days; see every CBK device

Record

Type
Abbreviated
Received
Oct 5, 2012
Review panel
Anesthesiology
Regulation
21 CFR 868.5895

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files

Logos are each company's own site icon, matched by name through Wikidata.Company names and logos identify the companies in public FDA and registry records and do not imply endorsement.