Arca
Skip to content

Modification To Servo-I Ventilator System, Model 64 87 800 E407E

Ventilator, Continuous, Facility Use. 510(k) K041223, from Maquet Critical Care AB (Bridgewater, NJ). Product code CBK, class 2.

As of Nov 17, 2026, K041223 is an FDA 510(k) held by Maquet Critical Care AB, substantially equivalent on Jul 29, 2004 after 80 days, with 8 recalls tied to it in FDA's recall database.

  • Substantially equivalent: Jul 29, 2004
  • Review time: 80 days
  • Recalls: 8
  • Cited as a predicate by: 10 devices

Indications for Use

As the 510(k) summary states them, read from FDA's summary PDF

The Servo¹ Ventilator System is intended for treatment and monitoring of patients in the range of neonates, infants and adults with respiratory failure or respiratory insufficiency. Servo¹ is a ventilator system to be used only by healthcare providers in hospitals or healthcare facilities and for in-hospital transport.

Predicate Devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a Predicate

10 later 510(k) summaries name this device as a predicate

Recalls

Recalls FDA ties to K041223, newest first; each opens FDA's recall record
Mar 7, 2013Z-1177-2013When using the SERVO-i on battery power, in a very small number of cases, some battery modules distributed after January 31, 2010 have a shorter battery run time than expected.Terminated
Dec 22, 2011Z-1362-2012The Servo-i and Servo-s ventilator systems equipped with specific revision os of the AC/DC convertor can result in electric shock during servicing.Terminated
Dec 22, 2011Z-1363-2012The Servo-i and Servo-s ventilator systems equipped with specific revision os of the AC/DC convertor can result in electric shock during servicing.Terminated
Sep 19, 2011Z-0352-2012Servo-I and Servo--S ventilator systems are equipped with a specific revision of a printed circuit board (PC1772 revision 22) that includes a component that has generated a number of nuisance alarms due to technical…Terminated
Sep 19, 2011Z-0353-2012Servo-I and Servo--S ventilator systems are equipped with a specific revision of a printed circuit board (PC1772 revision 22) that includes a component that has generated a number of nuisance alarms due to technical…Terminated
Sep 19, 2011Z-0354-2012Servo-I and Servo--S ventilator systems are equipped with a specific revision of a printed circuit board (PC1772 revision 22) that includes a component that has generated a number of nuisance alarms due to technical…Terminated
Jul 8, 2008Z-2442-2008Maquet became aware of potentially defective crimpings at the point connector attached to the cable in SERVO-i and SERVO-s ventilators.Terminated
Jul 8, 2008Z-2443-2008Maquet became aware of potentially defective crimpings at the point connector attached to the cable in SERVO-i and SERVO-s ventilators.Terminated

Other CBK Clearances

The newest 510(k)s for the same product code; their median review was 235 days; see every CBK device

Record

Type
Traditional
Received
May 10, 2004
Review panel
Anesthesiology
Regulation
21 CFR 868.5895

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files

Logos are each company's own site icon, matched by name through Wikidata.Company names and logos identify the companies in public FDA and registry records and do not imply endorsement.