Modification To Servo-I Ventilator System, Model 64 87 800 E407E
Ventilator, Continuous, Facility Use. 510(k) K041223, from Maquet Critical Care AB (Bridgewater, NJ). Product code CBK, class 2.
As of Nov 17, 2026, K041223 is an FDA 510(k) held by Maquet Critical Care AB, substantially equivalent on Jul 29, 2004 after 80 days, with 8 recalls tied to it in FDA's recall database.
- Substantially equivalent: Jul 29, 2004
- Review time: 80 days
- Recalls: 8
- Cited as a predicate by: 10 devices
Indications for Use
As the 510(k) summary states them, read from FDA's summary PDFThe Servo¹ Ventilator System is intended for treatment and monitoring of patients in the range of neonates, infants and adults with respiratory failure or respiratory insufficiency. Servo¹ is a ventilator system to be used only by healthcare providers in hospitals or healthcare facilities and for in-hospital transport.
Predicate Devices
The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF- Servo-I Ventilator System, Model 64 87 800 E407EK040221 · Maquet Critical Care AB · Feb 20, 2004
- Non-Invasive Ventilation Niv Option, Model 84 14 474K010093 · Drager Medizintechnik GmbH · Jun 14, 2001
Cited as a Predicate
10 later 510(k) summaries name this device as a predicate- Servo-I Ventilator SystemK123149 · Maquet Critical Care AB · Jun 20, 2014
- HAMILTON-C2K102775 · Hamilton Medical AG · Jan 21, 2011
- Engstrom Carestation, Engstrom ProK093886 · Datex-Ohmeda Inc., A Division of Instrumentarium C · May 3, 2010
- Engstrom Carestation, Engstrom ProK081842 · Datex-Ohmeda, Inc. · Oct 3, 2008
- Modification To HAMILTON-G5K081521 · Hamilton Medical AG · Sep 25, 2008
- Modification To Galileo GoldK080181 · Hamilton Medical AG · Jun 24, 2008
- Helontix VentK072926 · Linde Gas Therapeutics · Oct 15, 2008
- HAMILTON-G5K070513 · Hamilton Medical AG · Nov 26, 2007
- Servo-I Ventilator System, Models 64, 87 And 800K063404 · Maquet Critical Care AB · Feb 15, 2007
- Newport E500 Wave VentilatorK061094 · Newport Medical Instruments, Inc. · Jul 28, 2006
Recalls
Recalls FDA ties to K041223, newest first; each opens FDA's recall record| Mar 7, 2013 | Z-1177-2013When using the SERVO-i on battery power, in a very small number of cases, some battery modules distributed after January 31, 2010 have a shorter battery run time than expected. | Terminated |
| Dec 22, 2011 | Z-1362-2012The Servo-i and Servo-s ventilator systems equipped with specific revision os of the AC/DC convertor can result in electric shock during servicing. | Terminated |
| Dec 22, 2011 | Z-1363-2012The Servo-i and Servo-s ventilator systems equipped with specific revision os of the AC/DC convertor can result in electric shock during servicing. | Terminated |
| Sep 19, 2011 | Z-0352-2012Servo-I and Servo--S ventilator systems are equipped with a specific revision of a printed circuit board (PC1772 revision 22) that includes a component that has generated a number of nuisance alarms due to technical… | Terminated |
| Sep 19, 2011 | Z-0353-2012Servo-I and Servo--S ventilator systems are equipped with a specific revision of a printed circuit board (PC1772 revision 22) that includes a component that has generated a number of nuisance alarms due to technical… | Terminated |
| Sep 19, 2011 | Z-0354-2012Servo-I and Servo--S ventilator systems are equipped with a specific revision of a printed circuit board (PC1772 revision 22) that includes a component that has generated a number of nuisance alarms due to technical… | Terminated |
| Jul 8, 2008 | Z-2442-2008Maquet became aware of potentially defective crimpings at the point connector attached to the cable in SERVO-i and SERVO-s ventilators. | Terminated |
| Jul 8, 2008 | Z-2443-2008Maquet became aware of potentially defective crimpings at the point connector attached to the cable in SERVO-i and SERVO-s ventilators. | Terminated |
Other CBK Clearances
The newest 510(k)s for the same product code; their median review was 235 days; see every CBK device- Evita (V800); Evita (V600)K253647 · Draegerwerk AG & CO Kgaa · May 7, 2026
- Inbentus VersatileK252663 · Iconic Solutions by Murcia, SL · Apr 14, 2026
- VPAP Pediatric Face MaskK251888 · Hsiner Co., Ltd. · Mar 12, 2026
- Vivo 45 LSK232365 · Oconnell Regulatory Consultants, Inc. · Dec 19, 2025
- F&P OptiNIV Hospital Vented Full Face Mask Compatible with Single-limb Circuits Size A…K252173 · Fisher & Paykel Healthcare, Ltd. · Sep 10, 2025
- SV600, SV800 VentilatorK243767 · Shenzhen Mindray Bio-Medical Electronics Co., Ltd. · Aug 28, 2025
- AVEA disposable expiratory filter/water trap (11790)K251133 · Telesair, Inc. · May 8, 2025
- AF531 Oro-Nasal SE Face MaskK243394 · Respironics, Inc. · Dec 30, 2024
Record
- Document
- 510(k) summary (PDF) ↗
- Type
- Traditional
- Received
- May 10, 2004
- Review panel
- Anesthesiology
- Regulation
- 21 CFR 868.5895
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files