Arca
Skip to content
Tools
Tools

Carescape R860

Ventilator, Continuous, Facility Use. 510(k) K142679, from Datex-Ohmeda, Inc. (Madison, WI). Product code CBK, class 2.

As of Nov 20, 2026, K142679 is an FDA 510(k) held by Datex-Ohmeda, Inc., substantially equivalent on Jun 12, 2015 after 266 days, with 6 recalls tied to it in FDA's recall database.

  • Review time: 266 days
  • Recalls: 6
  • Cited as a predicate by: 2 devices

Indications for use

As the 510(k) summary states them, read from FDA's summary PDF

The CARESCAPE R860 ventilator is designed to provide mechanical ventilation or support to neonatal, pediatric, and adult patients weighing 0.25 kg and above. The CARESCAPE R860 ventilator is a microprocessor based, electronically controlled, pneumatically driven ventilator that includes integrated monitoring of FiO2, airway pressure, flow, and volume.

Predicate devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a predicate

2 later 510(k) summaries name this device as a predicate

Recalls

Recalls FDA ties to K142679, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
Apr 18, 2022Z-1226-2022, Z-1227-2022Insufficient battery backup power resulting in premature shutdown of the ventilator when not connected to AC mains power supply.Open
Jul 14, 2020Z-2883-2020The oxygen sensor for certain CARESCAPE R860 devices has a potential issue that may result in an inaccurate display of FiO2 value from what is being delivered.Terminated
Oct 5, 2018Z-0463-2019The ventilator Inspiratory Safety Guard (ISG) may disconnect from the breathing circuit pathway. As a result, this could create a loss of ventilation which may lead to inadequate oxygenation for patients, increasing the…Terminated
Jun 28, 2018Z-2738-2018The Neonatal Flow Sensor Cable may be missing the connector housing exposing the wires in the cable.Terminated
Dec 8, 2017Z-2266-2018Potential for the display processor to experience an unexpected failure.Terminated

Other CBK clearances

The newest 510(k)s for the same product code; their median review was 235 days; see every CBK device

Record

Type
Traditional
Received
Sep 19, 2014
Review panel
Anesthesiology
Regulation
21 CFR 868.5895

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files