Artistetm Solution With Sys Vcio
Accelerator, Linear, Medical. 510(k) K121295, from Siemens Medical Solutions USA, Inc. (Concord, CA). Product code IYE, class 2.
As of Nov 21, 2026, K121295 is an FDA 510(k) held by Siemens Medical Solutions USA, Inc., substantially equivalent on Jun 27, 2012 after 58 days, with 5 recalls tied to it in FDA's recall database.
- Review time: 58 days
- Recalls: 5
- Cited as a predicate by: 1 device
The intended use of the SIEMENS branded ARTISTET™, ONCOR™ and PRIMUST™ family of linear accelerator systems is to deliver X-ray photon and electron radiation for the therapeutic treatment of cancer. The linear accelerator systems are high-dose and high-dose rate medical linear accelerators optimized for 3D conformal radiation therapy, intensity-modulated radiation therapy (IMRT), modulated arc…
- Artiste Solution With SYS_VB50 UpdateK103606 · Siemens Medical Solutions USA, Inc. · Apr 15, 2011
- Artiste Solution, SYS_VC10C, Phase 2 Update With The Syngo RT Therapist & RT Oncologist Workspaces, V4.3.1_MR1K142434 · Siemens Medical Solutions USA, Inc. · Nov 21, 2014
| Feb 18, 2020 | Z-1677-2020Potential safety issue with the dose monitoring system safety interlocks which could result in a 10-20% overdose of radiation therapy. | Terminated |
| Mar 20, 2019 | Z-1204-2019Advisory issued not to disable or bypass the AMP function. Disabling or bypassing the AMP function for reasons of comfort is not advisable may lead to patient or user injury due to collision with LINAC equipment… | Terminated |
| Feb 14, 2018 | Z-1401-2018Control Console software has been updated to reduce the risk for collision when delivering automatically sequenced treatments with automatic movements of the gantry and/or the treatment table. | Terminated |
| May 29, 2013 | Z-1562-2013The potential for a safety issue, for the image quality degradation, that is addressed by the Customer Safety Advisory Notice when verifying the patient position based on 2D images that were acquired as single-exposure… | Terminated |
| Jan 17, 2013 | Z-0907-2013Siemens became aware of a potential safety issue when using the "In-session resumption" feature of the syngo RT Therapist (RTT) 4.3 software. The RTT will send again the previously completed beam or segment for delivery… | Terminated |
- AerFrame™ Alignment Laser Box (LBV2)K261865 · Stabilix, LLC · Sep 16, 2026
- EZFluence (RADEZ V2)K261718 · Radformation, Inc. · Aug 31, 2026
- MRIA Radiation Therapy Information System (MRIA)K260733 · Manteia Technologies Co., Ltd. · Aug 26, 2026
- Halcyon, Ethos Radiotherapy System (5.0)K261856 · Varian Medical Systems, Inc. · Aug 3, 2026
- ThinkQA 3K252937 · Dosisoft SA · Jun 12, 2026
- Luna 3DK253004 · Lap GmbH Laser Applikationen · Jun 4, 2026
- AerFrame Patient Immobilization System (AFPIS)K260792 · Stabilix, LLC · May 8, 2026
- Veriqa RT Epid 3DK252258 · Ptw-Freiburg Physikalisch-Technische-Werkstaetten Dr. Pychla · Apr 15, 2026
- Document
- 510(k) summary (PDF) ↗
- Type
- Traditional
- Received
- Apr 30, 2012
- Review panel
- Radiology
- Regulation
- 21 CFR 892.5050
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files