Artiste Solution With SYS_VB50 Update
Accelerator, Linear, Medical. 510(k) K103606, from Siemens Medical Solutions USA, Inc. (Concord, CA). Product code IYE, class 2.
As of Nov 20, 2026, K103606 is an FDA 510(k) held by Siemens Medical Solutions USA, Inc., substantially equivalent on Apr 15, 2011 after 127 days, with 7 recalls tied to it in FDA's recall database.
- Review time: 127 days
- Recalls: 7
- Cited as a predicate by: 1 device
The intended use of the SIEMENS branded ARTISTETM, ONCORTM and PRIMUSTM family of linear accelerator systems is to deliver X-ray photon and electron radiation for the therapeutic treatment of cancer. The linear accelerator systems are high-dose and high-dose rate medical linear accelerators optimized for 3D conformal radiation therapy, intensity-modulated radiation therapy (IMRT) and precision…
- Syngo RT Therapist Connect Workspace, V4.2K102671 · Siemens Medical Solutions USA, Inc. · Oct 12, 2010
- Artiste MV SAK072485 · Siemens Medical Solutions USA, Inc. · Dec 27, 2007
- Prescision OptionK082775 · Siemens Medical Solutions USA, Inc. · Feb 19, 2009
- Artistetm Solution With Sys VcioK121295 · Siemens Medical Solutions USA, Inc. · Jun 27, 2012
| Feb 18, 2020 | Z-1677-2020Potential safety issue with the dose monitoring system safety interlocks which could result in a 10-20% overdose of radiation therapy. | Terminated |
| Jul 26, 2016 | Z-2876-2016Software patch installation to address several safety issues. | Terminated |
| Aug 5, 2015 | Z-2812-2015A software fix has been released to prevent automatic movement resulting in a collision safety risk for patients. | Terminated |
| Apr 4, 2013 | Z-1182-2013Software update to fix multiple safety related issues. | Terminated |
| Jun 25, 2012 | Z-2073-2012Potential safety issue when using the IMMEDIATE RESUMPTION feature after a Control Console interlock during the delivery of flattening-filter-free (Multiple-X) beams or segments. In a very rare scenario, the RT… | Terminated |
| Apr 3, 2012 | Z-1486-2012SIEMENS has received reports where a dropping down of the 550 TxT" Treatment Table was observed while the table was being moved downwards during operation by means of the hand control. | Terminated |
| Nov 2, 2011 | Z-0462-2012The software release will bring current Installed Base system to the forward production version. It is a result of the final system testing addressing several safety issues that are not critical for the safe use of the… | Terminated |
- AerFrame™ Alignment Laser Box (LBV2)K261865 · Stabilix, LLC · Sep 16, 2026
- EZFluence (RADEZ V2)K261718 · Radformation, Inc. · Aug 31, 2026
- MRIA Radiation Therapy Information System (MRIA)K260733 · Manteia Technologies Co., Ltd. · Aug 26, 2026
- Halcyon, Ethos Radiotherapy System (5.0)K261856 · Varian Medical Systems, Inc. · Aug 3, 2026
- ThinkQA 3K252937 · Dosisoft SA · Jun 12, 2026
- Luna 3DK253004 · Lap GmbH Laser Applikationen · Jun 4, 2026
- AerFrame Patient Immobilization System (AFPIS)K260792 · Stabilix, LLC · May 8, 2026
- Veriqa RT Epid 3DK252258 · Ptw-Freiburg Physikalisch-Technische-Werkstaetten Dr. Pychla · Apr 15, 2026
- Document
- 510(k) summary (PDF) ↗
- Type
- Traditional
- Received
- Dec 9, 2010
- Review panel
- Radiology
- Regulation
- 21 CFR 892.5050
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files