Accelerator, Linear, Medical. 510(k) K103606, from Siemens Medical Solutions USA, Inc. (Concord, CA). Product code IYE, class 2.

As of Nov 20, 2026, K103606 is an FDA 510(k) held by Siemens Medical Solutions USA, Inc., substantially equivalent on Apr 15, 2011 after 127 days, with 7 recalls tied to it in FDA's recall database.

  • Review time: 127 days
  • Recalls: 7
  • Cited as a predicate by: 1 device

Indications for use

As the 510(k) summary states them, read from FDA's summary PDF

The intended use of the SIEMENS branded ARTISTETM, ONCORTM and PRIMUSTM family of linear accelerator systems is to deliver X-ray photon and electron radiation for the therapeutic treatment of cancer. The linear accelerator systems are high-dose and high-dose rate medical linear accelerators optimized for 3D conformal radiation therapy, intensity-modulated radiation therapy (IMRT) and precision…

Predicate devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a predicate

1 later 510(k) summary names this device as a predicate

Recalls

Recalls FDA ties to K103606, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
Feb 18, 2020Z-1677-2020Potential safety issue with the dose monitoring system safety interlocks which could result in a 10-20% overdose of radiation therapy.Terminated
Jul 26, 2016Z-2876-2016Software patch installation to address several safety issues.Terminated
Aug 5, 2015Z-2812-2015A software fix has been released to prevent automatic movement resulting in a collision safety risk for patients.Terminated
Apr 4, 2013Z-1182-2013Software update to fix multiple safety related issues.Terminated
Jun 25, 2012Z-2073-2012Potential safety issue when using the IMMEDIATE RESUMPTION feature after a Control Console interlock during the delivery of flattening-filter-free (Multiple-X) beams or segments. In a very rare scenario, the RT…Terminated
Apr 3, 2012Z-1486-2012SIEMENS has received reports where a dropping down of the 550 TxT" Treatment Table was observed while the table was being moved downwards during operation by means of the hand control.Terminated
Nov 2, 2011Z-0462-2012The software release will bring current Installed Base system to the forward production version. It is a result of the final system testing addressing several safety issues that are not critical for the safe use of the…Terminated

Other IYE clearances

The newest 510(k)s for the same product code; their median review was 173 days; see every IYE device

Record

Type
Traditional
Received
Dec 9, 2010
Review panel
Radiology
Regulation
21 CFR 892.5050

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files