Artiste MV SA
Accelerator, Linear, Medical. 510(k) K072485, from Siemens Medical Solutions USA, Inc. (San Diego, CA). Product code IYE, class 2.
As of Nov 20, 2026, K072485 is an FDA 510(k) held by Siemens Medical Solutions USA, Inc., substantially equivalent on Dec 27, 2007 after 114 days, with 24 recalls tied to it in FDA's recall database.
- Review time: 114 days
- Recalls: 24
- Cited as a predicate by: 6 devices
The intended use of the ARTÍSTE MV SAT TM linear accelerator system is to deliver x-ray radiation for therapeutic treatment of cancer. The ARTÍSTE MV SA™ includes an Electronic Portal Imaging Device (EPID), a 160 leaf multi-leaf collimator, and the syngo™ RT Therapist Express Workspace with MVision™.
- Oncor ExpressionK060226 · Siemens Medical Solutions USA, Inc. · Mar 15, 2006
- Artiste Solution With SYS_VB50 UpdateK103606 · Siemens Medical Solutions USA, Inc. · Apr 15, 2011
- Syngo RT Therapist Connect Workspace, V4.2K102671 · Siemens Medical Solutions USA, Inc. · Oct 12, 2010
- Linac Collimator Assembly Housing (Lcah) Model LcahartK100096 · Integra Radionics, Inc. · Apr 22, 2010
- 160MLC OptionK092145 · Siemens Medical Solutions USA, Inc. · Sep 18, 2009
- Syngo RT Therapist ConnectK090683 · Siemens Medical Solutions USA, Inc. · May 8, 2009
- Prescision OptionK082775 · Siemens Medical Solutions USA, Inc. · Feb 19, 2009
| Feb 18, 2020 | Z-1677-2020Potential safety issue with the dose monitoring system safety interlocks which could result in a 10-20% overdose of radiation therapy. | Terminated |
| Mar 20, 2019 | Z-1204-2019Advisory issued not to disable or bypass the AMP function. Disabling or bypassing the AMP function for reasons of comfort is not advisable may lead to patient or user injury due to collision with LINAC equipment… | Terminated |
| Feb 14, 2017 | Z-1488-2017Software update | Terminated |
| Jan 25, 2017 | Z-1212-2017Siemens became aware of Incorrect values for the rotational collimator position displayed due to mechanical problems of the sensor slide. In addition, there are improper weld seam at the overhead suspension. | Terminated |
| Dec 19, 2014 | Z-1059-2015A hardware update to replace the old PCB Assembly, Head Driver-G41 with PCB with overcurrent protection as a result of reports of the electric board inside the gantry showing signs of overheating, including thermal… | Terminated |
| Dec 2, 2014 | Z-0980-2015Use of any unauthorized third-party components on Siemens LINACs may lead to patient or operator injury, damage to the LINAC, and can potentially compromise customer warranty. The dark current dose must be considered in… | Terminated |
| Jun 17, 2014 | Z-2180-2014Siemens became aware of a potential issue with the syngo RT Therapist 4.3.SP1 and syngo RT Oncologist 4.3SP1 software applications running on Siemens brand LINAC systems that may experience sporadic system crashes that… | Terminated |
| May 28, 2014 | Z-2174-2014A safety risk exists with automatically sequenced treatment technique using the SIMTEC Auto Field Option where automatic movements of the gantry and the tabletop during an auto-sequenced treatment have the potential to… | Terminated |
| Apr 1, 2014 | Z-1645-2014Table may lose calibration during patient treatment, which may cause the possibility of mistreatment. | Terminated |
| Apr 1, 2014 | Z-1652-2014Siemens Radiation Oncology became aware that customers may be using the Siemens Linear Accelerator in combination with stereotactic accessories which have not been validated as being compatible with Siemens LINAC models. | Terminated |
| Oct 23, 2013 | Z-0438-2014A safety risk regarding a potentially existing dark current radiation phenomenon on the Linear Accelerator (LINAC) in combination with IMRT or mARC (rotational IMRT) treatments using unflat (flattening filter free)… | Terminated |
| Jan 22, 2012 | Z-1092-2012Software upgrade to installed base to the latest release of the RT Therapist version 4.1.144 which corrects multiple safety issues that could lead to patient mistreatment. | Terminated |
| Jun 3, 2011 | Z-3003-2011A potential safety issue was discovered when a pause for an imaging segment is inserted (where an imaging segment does not exist) with the ARTISTE and RT Therapist 4.1 system will continue to the next image. | Terminated |
| Dec 3, 2010 | Z-1422-2011The firm become aware of a potential issue with ARTISTE system that may result in patient injury should the positioner belt break, and the cam disc (secondary safety mechanism) is not installed and if the flat panel is… | Terminated |
| Jun 29, 2010 | Z-1523-2011A software problem that has the potential for data loss during the transfer of treatment records to the OIS, which may not be recorded in MOSAIQ, and subsequently may lead to mistreatment. | Terminated |
| Jan 22, 2010 | Z-0449-2011The robotic lift armature shaft may break and allow the Flat Panel Positioner with the digital flat panel imager to fall into the Flat Panel cradle. If the positioner is over a patient, there is a potential for injury. | Terminated |
| Sep 10, 2009 | Z-0121-2010A failed data transfer is possible that may result in data not being recorded in the interfaced system. Siemens Medical Solutions, Inc. USA became aware of this potential issue with its ARTISTE Linear Accelerator's… | Terminated |
| Aug 21, 2009 | Z-1970-2011When editing the note type of a previously entered note, then reverting to the original type, data may be lost. | Terminated |
| Jul 30, 2009 | Z-1064-2011Potential for Mistreatment-- when treatment plans intended for non-ARTISTE machinese are loaded, if the plans contain unspecified beam parameters, attributes or unknown block codes, the therapy system may remove those… | Terminated |
| Apr 9, 2009 | Z-1888-2009Unexpected treatment plan alteration-- Software issue may result in unintended modifications to treatment | Terminated |
| Dec 20, 2008 | Z-0949-2009The range of tolerance is to high for the accessory holder, therefore, Wedges and other accessories do not latch to the locking mechanism correctly. | Terminated |
| Oct 23, 2008 | Z-0857-2009Under certain conditions, the flat panel may move unexpectedly and collide with the patient. | Terminated |
| Oct 23, 2008 | Z-0927-2009If a plan contains a sequence of beams, including at least 1 ARC treatment beam and if during the setup stage of a beam following an ARC beam in the delivery sequence a termination occurs, and the user selects the… | Terminated |
| Jul 4, 2008 | Z-0116-2009Three issues: unexpected rotation of gantry, unexpected movement of table between beams during patient setup, and single exposure images will be overwritten by double exposure images. | Terminated |
- AerFrame™ Alignment Laser Box (LBV2)K261865 · Stabilix, LLC · Sep 16, 2026
- EZFluence (RADEZ V2)K261718 · Radformation, Inc. · Aug 31, 2026
- MRIA Radiation Therapy Information System (MRIA)K260733 · Manteia Technologies Co., Ltd. · Aug 26, 2026
- Halcyon, Ethos Radiotherapy System (5.0)K261856 · Varian Medical Systems, Inc. · Aug 3, 2026
- ThinkQA 3K252937 · Dosisoft SA · Jun 12, 2026
- Luna 3DK253004 · Lap GmbH Laser Applikationen · Jun 4, 2026
- AerFrame Patient Immobilization System (AFPIS)K260792 · Stabilix, LLC · May 8, 2026
- Veriqa RT Epid 3DK252258 · Ptw-Freiburg Physikalisch-Technische-Werkstaetten Dr. Pychla · Apr 15, 2026
- Document
- 510(k) summary (PDF) ↗
- Type
- Traditional
- Received
- Sep 4, 2007
- Review panel
- Radiology
- Regulation
- 21 CFR 892.5050
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files