Artiste Solution, SYS_VC10C, Phase 2 Update With The Syngo RT Therapist & RT Oncologist Workspaces, V4.3.1_MR1
Accelerator, Linear, Medical. 510(k) K142434, from Siemens Medical Solutions USA, Inc. (Martinez, CA). Product code IYE, class 2.
As of Nov 21, 2026, K142434 is an FDA 510(k) held by Siemens Medical Solutions USA, Inc., substantially equivalent on Nov 21, 2014 after 84 days, with 5 recalls tied to it in FDA's recall database.
- Review time: 84 days
- Recalls: 5
- Cited as a predicate by: 1 device
The intended use of the SIEMENS branded ARTISTETM, ONCOR™ and PRIMUSTM family of linear accelerator systems is to deliver X-Ray photon and electron radiation for the therapeutic treatment of cancer. The linear accelerator systems are high-dose and high-dose rate medical linear accelerators optimized for 3D conformal radiation therapy, intensity-modulated radiation therapy (IMRT), modulated arc…
- Artiste Solution With SYS_VC10C and Syngo RT Therapist Workspace, V4.3.1 Upgrade Kit, Aria Connectivity Option for SiemeK132935 · Siemens Medical Solutions USA, Inc. · Feb 21, 2014
- Artistetm Solution With Sys VcioK121295 · Siemens Medical Solutions USA, Inc. · Jun 27, 2012
- Linear Accelerator Control Console Software v13.1.010 Update, syngo Radiation Therapy Therapist v4.3.1_MR3 Patch, syngo…K182050 · Siemens Medical Solutions USA, Inc. · Oct 3, 2018
| Nov 9, 2022 | Z-0472-2023When selecting the site fraction group for treatment, the User may select the wrong site for treatment, and potentially deliver the dose to wrong isocenter | Open |
| Dec 9, 2021 | Z-0441-2022Potential safety issue under specific preconditons that may result in a user selecting the wrong site for treatment with the possibility to deliver dose to the wrong isocenter which could result in serious patient injury | Terminated |
| Aug 11, 2015 | Z-0106-2016This update is intended to provide a software update for the Syngo RT Oncologist, which is currently running SW version 4.2 or 4.3. The safety-related issues, which were described in the Customer Safety Notice… | Terminated |
| May 5, 2015 | Z-0168-2016potential safety risk of collision of the LINAC gantry with the patient or the treatment table in case of automatically sequenced treatment delivery techniques using SIMTEC" Auto Field Sequence Option. | Terminated |
| Jan 2, 2015 | Z-1013-2015Combination of CTVision with syngo RT Therapist / syngo RT Oncologist 4.3.SP1 automatic registration in Adaptive Targeting might result in wrong offset calculations. Applying this offset can lead to patient… | Terminated |
- AerFrame™ Alignment Laser Box (LBV2)K261865 · Stabilix, LLC · Sep 16, 2026
- EZFluence (RADEZ V2)K261718 · Radformation, Inc. · Aug 31, 2026
- MRIA Radiation Therapy Information System (MRIA)K260733 · Manteia Technologies Co., Ltd. · Aug 26, 2026
- Halcyon, Ethos Radiotherapy System (5.0)K261856 · Varian Medical Systems, Inc. · Aug 3, 2026
- ThinkQA 3K252937 · Dosisoft SA · Jun 12, 2026
- Luna 3DK253004 · Lap GmbH Laser Applikationen · Jun 4, 2026
- AerFrame Patient Immobilization System (AFPIS)K260792 · Stabilix, LLC · May 8, 2026
- Veriqa RT Epid 3DK252258 · Ptw-Freiburg Physikalisch-Technische-Werkstaetten Dr. Pychla · Apr 15, 2026
- Document
- 510(k) summary (PDF) ↗
- Type
- Traditional
- Received
- Aug 29, 2014
- Review panel
- Radiology
- Regulation
- 21 CFR 892.5050
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files