Arca
Skip to content
Tools
Tools

Mobilett Mira

System, X-Ray, Mobile. 510(k) K111912, from Siemens Medical Solutions USA, Inc. (Mavern, PA). Product code IZL, class 2.

As of Nov 21, 2026, K111912 is an FDA 510(k) held by Siemens Medical Solutions USA, Inc., substantially equivalent on Sep 19, 2011 after 76 days, with 5 recalls tied to it in FDA's recall database.

  • Review time: 76 days
  • Recalls: 5
  • Cited as a predicate by: 2 devices

Indications for use

As the 510(k) summary states them, read from FDA's summary PDF

The Mobilett Mira is a radiographic system designed for use in wards, intensive care and premature-birth wards, pediatric and emergency departments, operating theatres as well as the central X-ray department.

Predicate devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a predicate

2 later 510(k) summaries name this device as a predicate

Recalls

Recalls FDA ties to K111912, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
Oct 5, 2023Z-0223-2024Due to the incorrect mounting of the buzzer component, there is no acoustic signal during x ray activation. Otherwise, the system is fully functional, and x ray is applied correctly. By means of the additional visual…Open
May 2, 2013Z-1875-2013Siemens issued a Field Safety Alert about the potential risk to users when operating the Mobilett Mira. When a Mobilett Mira is regularly exposed to extensive mechanical strain (driven over high thresholds for doors or…Terminated
Oct 10, 2012Z-0480-2013There is a risk of lost images during examination.Terminated
Aug 21, 2012Z-2322-2012During regular product monitoring, Siemens has become aware of a potential issue with the Mobilett Mira system. During normal clinical operation the switch on the motor drive handle of the unit may become loose and…Terminated
Apr 30, 2012Z-2367-2012During regular product monitoring, firm became aware of a potential issue with the Mobilett Mira system, (serial numbers 1034, 1038, 1041, 1042, 1043, 1050, 1053, and 1054. During normal clinical operation the center…Terminated

Other IZL clearances

The newest 510(k)s for the same product code; their median review was 166 days; see every IZL device

Record

Type
Special
Received
Jul 5, 2011
Review panel
Radiology
Regulation
21 CFR 892.1720

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files