Arca
Skip to content
Tools
Tools

Ysio

System, X-Ray, Stationary. 510(k) K081722, from Siemens Medical Solutions USA, Inc. (Malvern, PA). Product code KPR, class 2.

As of Nov 19, 2026, K081722 is an FDA 510(k) held by Siemens Medical Solutions USA, Inc., substantially equivalent on Aug 22, 2008 after 65 days, with 11 recalls tied to it in FDA's recall database.

  • Review time: 65 days
  • Recalls: 11
  • Cited as a predicate by: 7 devices

Indications for use

As the 510(k) summary states them, read from FDA's summary PDF

The Ysio (New RAD –FAMILY) systems are radiographic systems used in hospitals, clinics, and medical practices. Ysio enables radiographic and tomographic exposures of the whole body including: skull, chest, abdomen, and extremities and may be used on pediatric, adult and bariatric patients.

Predicate devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a predicate

7 later 510(k) summaries name this device as a predicate

Recalls

Recalls FDA ties to K081722, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
Jan 31, 2022Z-0841-2022Siemens Healthineers has identified a software error in previous software version VB10 for the Ysio fully automated system. The unit may process incorrect internal values for the position of its ceiling tube stand. If…Open
Feb 10, 2020Z-1501-2020The Ysio system may process incorrect internal values for the position of its ceiling tube stand. If in this situation the user activates longitudinal or transversal stand movement (by using the buttons at the tube head…Terminated
Aug 6, 2018Z-0299-2019In rare cases of high clinical workload, the steel cables inside the lifting column of the overhead tube, which are designed to take the load, may rapture without triggering the safety lock. This could lead to…Terminated
Aug 27, 2014Z-2680-2014Siemens discovered that an unlikely error may occur on the Ysio system with fixed detector in the wall stand. This error may result in line artifacts in the image. If these artifacts appear in the region of interest,…Terminated
Jul 7, 2014Z-2292-2014It was discovered that during a RAD examination using Siemens Luminos dRF, Ysio or Uroskop Omnia systems with software version VB10C to VB10F and automatic exposure control, a highly unlikely malfunction can result in…Terminated
May 9, 2014Z-1717-2014Excessive usage of cleaning liquid on Siemens Ysio Systems may cause infiltration of the generator console, which may lead to a short circuit causing an unintended release of radiation.Terminated
Sep 3, 2013Z-0374-2014A potential malfunction and hazard to patients exists when using the Ysio, Luminos Agile, Luminos dRF, and Uroskop Omnia systems with software version VC10, which may result in radiation continuing longer than required…Terminated
Aug 28, 2013Z-1845-2014When using systems operating with software versions VD10A/G during a RAD examination, a malfunction can result in the image appearing very dark, which may prompt unnecessary repetition of the examination.Terminated
May 6, 2013Z-0016-2014Siemens has discovered through product monitoring a potential malfunction when using system operating with software versions VB10D, VB10F, VB10C and VB10. During a RAD examination the possibility exists that radiation…Terminated
Feb 15, 2012Z-1159-2012Firm has become aware of unintended behavior when using Ysio, AXIOM dRF, AXIOM Luminos Agile/TF and Uroskop Omnia with RAD Fluoro Uro function using software version VC10A. Due to an error in the image pipeline, images…Terminated
Apr 28, 2011Z-2787-2011There is a potential hazard to staff or patients when lowering the table height on the Ysio system. If staff or patients lower extremities are placed underneath the table and the user lowers the table top, it is…Terminated

Other KPR clearances

The newest 510(k)s for the same product code; their median review was 195 days; see every KPR device

Record

Type
Traditional
Received
Jun 18, 2008
Review panel
Radiology
Regulation
21 CFR 892.1680

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files