Ysio
System, X-Ray, Stationary. 510(k) K081722, from Siemens Medical Solutions USA, Inc. (Malvern, PA). Product code KPR, class 2.
As of Nov 19, 2026, K081722 is an FDA 510(k) held by Siemens Medical Solutions USA, Inc., substantially equivalent on Aug 22, 2008 after 65 days, with 11 recalls tied to it in FDA's recall database.
- Review time: 65 days
- Recalls: 11
- Cited as a predicate by: 7 devices
The Ysio (New RAD –FAMILY) systems are radiographic systems used in hospitals, clinics, and medical practices. Ysio enables radiographic and tomographic exposures of the whole body including: skull, chest, abdomen, and extremities and may be used on pediatric, adult and bariatric patients.
- Thorax FD and Multix FDK983732 · Siemens Medical Solutions USA, Inc. · Apr 21, 1999
- Axiom Aristos FX PlusK061054 · Siemens Medical Solutions USA, Inc. · May 9, 2006
- Krystal-X WifiK070505 · Owandy · Apr 5, 2007
- DigiX FDXK223060 · Allengers Medical Systems Limited · Oct 26, 2022
- General Radiography X-ray System (Model: GR10X 40K), General Radiography X-ray System (Model: GR10X 50K)K220149 · Xera Medical Systems & Technology , Ltd. · Apr 19, 2022
- DigiX FDXK192541 · Allengers Medical Systems Limited · Oct 16, 2019
- DigiX FDXK162529 · Allengers Medical Systems Limited · Mar 3, 2017
- Ysio MaxK133259 · Siemens Medical Solutions USA, Inc. · Jan 24, 2014
- Multix FusionK121513 · Siemens Medical Solution · Aug 10, 2012
- Mobilett MiraK111912 · Siemens Medical Solutions USA, Inc. · Sep 19, 2011
| Jan 31, 2022 | Z-0841-2022Siemens Healthineers has identified a software error in previous software version VB10 for the Ysio fully automated system. The unit may process incorrect internal values for the position of its ceiling tube stand. If… | Open |
| Feb 10, 2020 | Z-1501-2020The Ysio system may process incorrect internal values for the position of its ceiling tube stand. If in this situation the user activates longitudinal or transversal stand movement (by using the buttons at the tube head… | Terminated |
| Aug 6, 2018 | Z-0299-2019In rare cases of high clinical workload, the steel cables inside the lifting column of the overhead tube, which are designed to take the load, may rapture without triggering the safety lock. This could lead to… | Terminated |
| Aug 27, 2014 | Z-2680-2014Siemens discovered that an unlikely error may occur on the Ysio system with fixed detector in the wall stand. This error may result in line artifacts in the image. If these artifacts appear in the region of interest,… | Terminated |
| Jul 7, 2014 | Z-2292-2014It was discovered that during a RAD examination using Siemens Luminos dRF, Ysio or Uroskop Omnia systems with software version VB10C to VB10F and automatic exposure control, a highly unlikely malfunction can result in… | Terminated |
| May 9, 2014 | Z-1717-2014Excessive usage of cleaning liquid on Siemens Ysio Systems may cause infiltration of the generator console, which may lead to a short circuit causing an unintended release of radiation. | Terminated |
| Sep 3, 2013 | Z-0374-2014A potential malfunction and hazard to patients exists when using the Ysio, Luminos Agile, Luminos dRF, and Uroskop Omnia systems with software version VC10, which may result in radiation continuing longer than required… | Terminated |
| Aug 28, 2013 | Z-1845-2014When using systems operating with software versions VD10A/G during a RAD examination, a malfunction can result in the image appearing very dark, which may prompt unnecessary repetition of the examination. | Terminated |
| May 6, 2013 | Z-0016-2014Siemens has discovered through product monitoring a potential malfunction when using system operating with software versions VB10D, VB10F, VB10C and VB10. During a RAD examination the possibility exists that radiation… | Terminated |
| Feb 15, 2012 | Z-1159-2012Firm has become aware of unintended behavior when using Ysio, AXIOM dRF, AXIOM Luminos Agile/TF and Uroskop Omnia with RAD Fluoro Uro function using software version VC10A. Due to an error in the image pipeline, images… | Terminated |
| Apr 28, 2011 | Z-2787-2011There is a potential hazard to staff or patients when lowering the table height on the Ysio system. If staff or patients lower extremities are placed underneath the table and the user lowers the table top, it is… | Terminated |
- EXSYS DEXi (EXSYS DEXi-S40S); EXSYS DEXi (EXSYS DEXi-S40P); EXSYS DEXi (EXSYS DEXi-S40T); EXSYS DEXi (EXSYS DEXi-S50HK260227 · DRTECH Corporation · Sep 8, 2026
- MULTIX Impact; MULTIX Impact CK260488 · Siemens Medical Solutions USA, Inc. · Aug 31, 2026
- GC85AK260481 · Samsung Electronics Co., Ltd. · Aug 27, 2026
- DRE DuoK260482 · Jpi Healthcare Co, Ltd. · Jun 15, 2026
- AeroDR TX c02K253446 · Dk Medical Systems Co., Ltd. · Apr 10, 2026
- DRX-Evolution Plus X-ray System, DRX-Compass Mobile X-ray System, Lux HD 35 detector, Lux HD 43 detectorK250954 · Carestream Health · Jan 20, 2026
- uDR Arria & uDR ArisK252000 · Shanghai United Imaging Healthcare Co., Ltd. · Nov 26, 2025
- uDR Aurora CXK251167 · Shanghai United Imaging Healthcare Co., Ltd. · Sep 19, 2025
- Document
- 510(k) summary (PDF) ↗
- Type
- Traditional
- Received
- Jun 18, 2008
- Review panel
- Radiology
- Regulation
- 21 CFR 892.1680
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files