Arca
Skip to content
Tools
Tools

Mobilette XP and Mobilett Xphybrid

System, X-Ray, Mobile. 510(k) K033238, from Siemens Medical Systems, Inc. (Mavern, PA). Product code IZL, class 2.

As of Nov 21, 2026, K033238 is an FDA 510(k) held by Siemens Medical Systems, Inc., substantially equivalent on Nov 14, 2003 after 39 days, with 4 recalls tied to it in FDA's recall database.

  • Review time: 39 days
  • Recalls: 4
  • Cited as a predicate by: 1 device

Indications for use

As the 510(k) summary states them, read from FDA's summary PDF

The Mobilett XP and Mobilett XP$_{hybrid}$ are radiographic systems designed for use in wards, intensive care and premature-birth wards, pediatric and emergency departments, operating theatres as well as the central X-ray department.

Predicate devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a predicate

1 later 510(k) summary names this device as a predicate
  • Mobilett MiraK111912 · Siemens Medical Solutions USA, Inc. · Sep 19, 2011

Recalls

Recalls FDA ties to K033238, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
May 9, 2014Z-1985-2014There is a potential fire hazard of certain Lithium-Ion batteries in the computers used in the original manufacturing or as spare parts between October 2010 and April 2011 for the Siemens Mobilett XP Digital mobile…Terminated
Oct 4, 2010Z-0171-2011Units not equipped with protective plastic cover over electronics.Terminated
Oct 4, 2010Z-0718-2011battery fuseTerminated
Mar 21, 2007Z-0749-2007Mounting bolts for tank fork assembly and collimator flange may become loose.Terminated

Other IZL clearances

The newest 510(k)s for the same product code; their median review was 166 days; see every IZL device

Record

Type
Special
Received
Oct 6, 2003
Review panel
Radiology
Regulation
21 CFR 892.1720

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files

Esc

Type a name: a tool, a company, a drug or brand, a condition, a gene or a state. Identifiers like K231234, NCT05012345 or NDA 215866 open their record.

Logos are each company's own site icon, matched by name through Wikidata, or logos provided by Logo.dev.Company names and logos identify the companies in public FDA and registry records and do not imply endorsement.