Cadd Solis Vip Ambulatory Infusion Pump and Cadd Solis Medication Safety Software
Pump, Infusion. 510(k) K111275, from Smiths Medical Asd, Inc. (St. Paul, MN). Product code FRN, class 2.
As of Nov 20, 2026, K111275 is an FDA 510(k) held by Smiths Medical Asd, Inc., substantially equivalent on Feb 1, 2013 after 637 days, with 6 recalls tied to it in FDA's recall database.
- Review time: 637 days
- Recalls: 6
- Cited as a predicate by: 4 devices
"The CADD ® -Solis VIP Ambulatory Infusion Pump is indicated for intravenous, intra-arterial, subcutaneous, intraperitoneal, perineural, surgical site, epidural space, or subarachnoid space infusion. PCA (patient-controlled analgesia) delivery is used for therapies that require a continuous rate of infusion, patient-controlled demand doses, or both, such as patient-controlled analgesia.
- Cadd -Solis Ambulatory Infusion Pump and Cadd -Solis Medication Safety Software - AdministratorK072144 · Smiths Medical MD, Inc. · Mar 7, 2008
- Cadd-Prizm Model 6100 PumpK960826 · Sims Deltec, Inc. · Aug 23, 1996
- Cadd -Solis Medication Safety Software: Administrator and Point-Of-CareK082783 · Smiths Medical MD, Inc. · Dec 17, 2008
- Curlin 8000 Ambulatory Infusion System; Enterprise Solution (ES) Software; Curlin 380-Series Adminstration SetsK232793 · Zevex, Inc. · May 17, 2024
- Halo II Ambulatory Infusion System (also with trade names Nimbus II Ambulatory Infusion System, EVAA Ambulatory…K153193 · Zyno Medical, LLC · Aug 30, 2016
- Halo Ambulatory Infusion SystemK140783 · Zyno Medical, LLC · Oct 17, 2014
- Cadd -Solis Ambulatory Infusion Pump, Version 3.0K130394 · Smiths Medical Asd, Inc. · Dec 9, 2013
| Apr 10, 2025 | Z-1668-2025Affected pumps may trigger an erroneous (false) Upstream Occlusion Alarm under certain conditions, which will interrupt an active infusion. Interruption or delay of therapy can lead to serious patient injury or death. | Open |
| Apr 10, 2025 | Z-1682-2025There is a potential for thermal damage in CADD-Solis and CADD- Solis VIP infusion pumps. | Open |
| Jul 16, 2024 | Z-0222-2025Damage to the battery pack may have caused a short to a capacitor within the battery pack. While the battery encasement is designed to be flame retardant, a short to the capacitor could lead to melting of the battery… | Open |
| Jun 26, 2024 | Z-0466-2025The adapter input plug can become damaged or broken. If the input plug is damaged, the metal contacts to the body of the AC adapter may be exposed or one or more of the AC mains prongs may separate from the input plug. | Open |
| Feb 27, 2024 | Z-2421-2024The CADD Solis VIP Ambulatory Infusion Pump is indicated for the following uses: " For intravenous, intraarterial, subcutaneous, intraperitoneal, perineural, surgical site, epidural space, or subarachnoid space infusion. | Open |
| Mar 24, 2020 | Z-0107-2022Specific Pumps may exhibit intermittent performance in the AILD (Air in Line Detector) function. If an AILD does not sufficiently discern fluid from air in line, an air in line event may not be recognized by the pump… | Terminated |
- RemunityPRO+ Pump for Remodulin® (treprostinil) InjectionK253703 · Deka R&D Corp · Jul 24, 2026
- FreedomEdge Infusion System; High-Flo SubQ Needle Set; Precision Flow Rate TubingK252015 · Koru Medical Systems, Inc. · Jan 28, 2026
- Spectrum IQ Infusion System with Dose IQ Safety Software (3570009)K251636 · Baxter Healthcare Corporation · Jul 28, 2025
- SIGMA Spectrum Infusion Pump with Master Drug LibraryK251640 · Baxter Healthcare Corporation · Jul 28, 2025
- MRidium 3870 MRI Infusion Pump System (3870)K242752 · Iradimed Corporation · May 28, 2025
- BD Alaris Infusion System with Guardrails Suite MXK243855 · Carefusion 303, Inc. · Apr 25, 2025
- Plum Solo™ Precision IV PumpK242114 · Icu Medical, Inc. · Apr 2, 2025
- Plum Duo™ Precision IV PumpK242115 · Icu Medical, Inc. · Apr 2, 2025
- Document
- 510(k) summary (PDF) ↗
- Type
- Traditional
- Received
- May 6, 2011
- Review panel
- General Hospital
- Regulation
- 21 CFR 880.5725
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files