Cadd -Solis Ambulatory Infusion Pump and Cadd -Solis Medication Safety Software - Administrator

Pump, Infusion, Pca. 510(k) K072144, from Smiths Medical MD, Inc. (St. Paul, MN). Product code MEA, class 2.

As of Nov 20, 2026, K072144 is an FDA 510(k) held by Smiths Medical MD, Inc., substantially equivalent on Mar 7, 2008 after 218 days, with 7 recalls tied to it in FDA's recall database.

  • Review time: 218 days
  • Recalls: 7
  • Cited as a predicate by: 2 devices

Indications for use

As the 510(k) summary states them, read from FDA's summary PDF

"The CADD®-Solis ambulatory infusion pump is indicated for intravenous, intra-arterial, subcutaneous, intraperitoneal, in close proximity to nerves, into an intraoperative site (soft tissue, body cavity/surgical wound site), epidural space, or subarachnoid space infusion.

Cited as a predicate

2 later 510(k) summaries name this device as a predicate

Recalls

Recalls FDA ties to K072144, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
Apr 10, 2025Z-1667-2025Affected pumps may trigger an erroneous (false) Upstream Occlusion Alarm under certain conditions, which will interrupt an active infusion. Interruption or delay of therapy can lead to serious patient injury or death.Open
Feb 27, 2024Z-2419-2024CADD-Solis Ambulatory Infusion Pumps, with software versions before v4.3, may have the following issues:1. Upstream Occlusion, 2. Stop and Power Keys Unresponsive, 3. Manual Mode Air Detector, 4. Single Bubble Air…Open
May 26, 2015Z-1813-2015Smiths Medical has become aware of an issue with an intermittent occurrence of binding of the locking assembly on some CADD-Solis pumps. Binding of the Cassette/ Keypad Lock can occur after latching the CADD-Solis…Terminated
Sep 29, 2014Z-0166-2015Smiths Medical has identified an issue with a single batch (Lot Number 2752712) of CADD"-Solis Medication Safety Software Administrator 3.1 CDs. CADD"-Solis Medication Safety Software Administrator 3.1 CDs were sent…Terminated
May 10, 2011Z-0004-2012Reports of non-delivery/ significant under-delivery of fluids when the CADD Medication Cassette Reservoir is not correctly latched to the pump. EXPANDED: On 7/14/11, Smith Medical expanded this recall to include all…Terminated
Jan 13, 2011Z-1033-2011Reports of the Cassette Lock mechanism on certain CADD-Solis Pumps becoming stuck in the locked position, making it nearly impossible to remove the medication cassette reservoirs (or administration set). The purpose of…Terminated
Jan 27, 2010Z-1082-2010Potential for the power cord's prongs to crack and fail at/or inside the plug. Other medical device manufacturers have reported incidents of sparking, charring, and fires from the affected power cords used with their…Terminated

Record

Type
Traditional
Received
Aug 2, 2007
Review panel
General Hospital
Regulation
21 CFR 880.5725

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files