Arca
Skip to content

Spacelabs Medical Patient Monitors

Monitor, Physiological, Patient(With Arrhythmia Detection Or Alarms). 510(k) K102422, from Spacelabs Medical, Inc. (Issaquah, WA). Product code MHX, class 2.

As of Nov 17, 2026, K102422 is an FDA 510(k) held by Spacelabs Medical, Inc., substantially equivalent on Dec 7, 2010 after 104 days, with 3 recalls tied to it in FDA's recall database.

  • Substantially equivalent: Dec 7, 2010
  • Review time: 104 days
  • Recalls: 3
  • Cited as a predicate by: 3 devices

Indications for Use

As the 510(k) summary states them, read from FDA's summary PDF

Spacelabs Healthcare patient monitors, functioning as either bedside or central monitors; passively display data generated by Spacelabs Healthcare parameter modules, Flexports interfaces, and other SDLC based products in the form of waveform and numeric displays, trends and alarms.

Predicate Devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a Predicate

3 later 510(k) summaries name this device as a predicate

Recalls

Recalls FDA ties to K102422, newest first; each opens FDA's recall record
Aug 26, 2014Z-2707-2014The clinician did not realize the monitor was in END CASE mode and was expecting audible alarms. Visual alarms continue to work normally. The firm performed a retrospective analysis of all recalls since 2011 and…Terminated
Jul 8, 2011Z-3032-2011The clinician did not realize the monitor was in END CASE mode and was expecting audible alarms. Visual alarms continue to work normally.Terminated
Mar 24, 2011Z-2134-2011The circuit that supplies backup power to the modules in the Medical Patient Monitor, Model 91387, is not working. If there is a power interruption, the three-minute backup may not be operational and customers may lose…Terminated

Other MHX Clearances

The newest 510(k)s for the same product code; their median review was 226 days; see every MHX device

Record

Type
Special
Received
Aug 25, 2010
Review panel
Cardiovascular
Regulation
21 CFR 870.1025

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files

Logos are each company's own site icon, matched by name through Wikidata.Company names and logos identify the companies in public FDA and registry records and do not imply endorsement.