Spacelabs Healthcare Qube Compact Monitor (91390
Monitor, Physiological, Patient(With Arrhythmia Detection Or Alarms). 510(k) K120616, from Spacelabs Medical, Inc. (Nederland, CO). Product code MHX, class 2.
As of Nov 17, 2026, K120616 is an FDA 510(k) held by Spacelabs Medical, Inc., substantially equivalent on Mar 29, 2012 after 29 days, with 9 recalls tied to it in FDA's recall database.
- Substantially equivalent: Mar 29, 2012
- Review time: 29 days
- Recalls: 9
Indications for Use
As the 510(k) summary states them, read from FDA's summary PDFSpacelabs Healthcare Qube Compact Monitor (91390) The Spacelabs Healthcare Qube Compact Monitor (91390), functioning as either bedside or central monitors; passively displays data generated by Spacelabs Healthcare parameter modules, Flexports interfaces, and other SDLC based products in the form of waveform and numeric displays, trends and alarms.
Predicate Devices
The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF- Spacelabs Medical Patient MonitorsK102422 · Spacelabs Medical, Inc. · Dec 7, 2010
Recalls
Recalls FDA ties to K120616, newest first; each opens FDA's recall record| Sep 5, 2025 | Z-0167-2026Due to display cables used for monitors were not in compliance with electromagnetic interference from inadequately shielded cables which may pose serious risk to patient's safety by interfering with the operation of… | Open, Classified |
| Sep 5, 2025 | Z-0169-2026Due to display cables used for monitors were not in compliance with electromagnetic interference from inadequately shielded cables which may pose serious risk to patient's safety by interfering with the operation of… | Open, Classified |
| Sep 5, 2025 | Z-0168-2026Due to display cables used for monitors were not in compliance with electromagnetic interference from inadequately shielded cables which may pose serious risk to patient's safety by interfering with the operation of… | Open, Classified |
| May 8, 2015 | Z-1608-2015The input power can fail even though the power source at the wall has not been interrupted. In addition, there is the potential for the qube Compact Monitor to fail to turn on. This is a retrospective recall related to… | Terminated |
| Jan 28, 2015 | Z-1145-2015Spacelabs Healthcare qube Bedside Monitors, Model 91390, are recalled because the firm has received multiple reports of qube monitors failing to boot up or returning to factory default configuration settings following… | Terminated |
| Jun 6, 2014 | Z-1852-2014Spacelabs Healthcare is recalling Qube Compact Monitor, Model 91390, due to a potential chance of losing network connection after switching from hardwired Ethernet connection to using the Qube Docking Station without… | Terminated |
| Oct 17, 2013 | Z-0351-2014Spacelabs Healthcare qube Compact Patient Monitor, Model 91390, may fail to power ON due to incorrect programmed microcontroller component. | Terminated |
| Oct 17, 2013 | Z-0188-2014Spacelabs Healthcare elance Vital Signs Monitor with Option S, model 93330, is recalled because the monitor will not permit connection of any of the Spacelabs TruLink line of SpO2 sensors. | Terminated |
| Oct 19, 2012 | Z-0403-2013Spacelabs Healthcare has learned through several reports that the input circuit may fail and the monitor will go on battery mode. The monitor will function normally until the batteries become depleted and then the… | Terminated |
Other MHX Clearances
The newest 510(k)s for the same product code; their median review was 226 days; see every MHX device- Radius VSM and AccessoriesK260676 · Masimo Corporation · Oct 14, 2026
- Carescape ONE; Carescape Dock F0; TruSignal ECG MM; TruSignal PRES MM; TruSignal TEMP MM; TruSignal SpO2 MM 1.8M;…K260656 · GE Healthcare Finland Oy · Sep 1, 2026
- Carescape Canvas 1000; Carescape Canvas Smart DisplayK260343 · GE Healthcare Finland Oy · Aug 28, 2026
- Infinity® Acute Care System (IACS) VG8.1K260579 · Draeger Medical Systems, Inc. · Aug 20, 2026
- Telemetry Monitor (iT30, iT50)K251508 · Edan Instruments, Inc. · Aug 6, 2026
- Monitor B105M; Monitor B125M; Monitor B155M; Monitor B105P; Monitor B125PK253277 · Ge Medical Systems Information Technologies, Inc. · Jul 24, 2026
- Radius VSM and AccessoriesK254209 · Masimo Corporation · Jun 4, 2026
- M12 Telemetry SystemK252840 · Global Instrumentation, LLC · May 29, 2026
Record
- Document
- 510(k) summary (PDF) ↗
- Type
- Special
- Received
- Feb 29, 2012
- Review panel
- Cardiovascular
- Regulation
- 21 CFR 870.1025
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files